DEN-7117: changed shock threshold from 5 to 2.5 g.

DEN-7117: Changed shock threshold from 5 to 2.5 g.

Merge branch 'DEN-6372-hd-rinseback' into develop

Merge branch 'staging' into DEN-6372-hd-rinseback

HD-DEN-6372_HD Rinseback
HD-DEN-6372_HD Rinseback
Merge branch 'staging' into DEN-6372-hd-rinseback

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed.

Fixed.

Need to fix doxygen style. @details Inputs: @details Outputs:

Need to fix doxygen style.
@details Inputs:
@details Outputs:

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed.

Removed.

I think setActiveReservoirCMD() sets activeReservoir.data for you. Redundant.

I think setActiveReservoirCMD() sets activeReservoir.data for you. Redundant.

Due to changes from non-volatile data management, this message is no longer available. Need to remove to fix build.

Due to changes from non-volatile data management, this message is no longer available.
Need to remove to fix build.

RTCs are sync'd by UI sending its date/time to HD and DG f/w. However, RTC time is not used for treatment time. Treatment time is accumulated using HD timer counter. HD then uses this and set treat...

RTCs are sync'd by UI sending its date/time to HD and DG f/w.
However, RTC time is not used for treatment time. Treatment time is accumulated using HD timer counter. HD then uses this and set treatment duration to calculate time remaining. All are then sent to UI regularly.

Done.

Done.

Why removed?

Why removed?

Aligned.

Aligned.

1. How are we syncing times between UI RTC and HD RTC? 2. Does not the UI APP know some of these times (like start time) to compute the other times by itself instead of HD FW passing them to the UI...

1. How are we syncing times between UI RTC and HD RTC?
2. Does not the UI APP know some of these times (like start time) to compute the other times by itself instead of HD FW passing them to the UI APP?

Align comment.

Align comment.