DENBUG-59: Addressed some compiler warnings.

Merge branch 'staging' into develop

DEN-15979 For testing

DEN-15979 For testing

DEN-15979 For testing

DEN-15714 In progress

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DENBUG-64

HD-DEN-15714_EE Alarm 84 HD Rtc Time After Cycling DG And HD Dvt 02
HD-DEN-15714_EE Alarm 84 HD Rtc Time After Cycling DG And HD Dvt 02
DENBUG64 removed the blood detect windowed timer

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.