fwcommon

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Local logging is a future requirement. Some work was done long ago before we knew it wasn't needed yet. #ifdef added to prevent compiler warnings about non-usage.

Local logging is a future requirement. Some work was done long ago before we knew it wasn't needed yet. #ifdef added to prevent compiler warnings about non-usage.

Should queueFrontIndex be added to header? Is ENABLE_LOCAL_LOGGING only for testing purposes?

Should queueFrontIndex be added to header? Is ENABLE_LOCAL_LOGGING only for testing purposes?

DG-DENBUG-59_FW Duplicate Alarms Occur During Post Treatment Rinseback And Block Workflow
DG-DENBUG-59_FW Duplicate Alarms Occur During Post Treatment Rinseback And Block Workflow
HD-DEN-15714_EE Alarm 84 HD Rtc Time After Cycling DG And HD Dvt 02
HD-DEN-15714_EE Alarm 84 HD Rtc Time After Cycling DG And HD Dvt 02
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Done

Done

Move up with other setup functions.

Move up with other setup functions.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Should be the recovery alarm.

Should be the recovery alarm.

HD-DENBUG-64_Sys Dialyzer Blood Leak Alarm Triggered During Treatment
HD-DENBUG-64_Sys Dialyzer Blood Leak Alarm Triggered During Treatment