DEN-5887: Moved some load cell functions from dialysate outlet pump controller to DG Interface module. Added small and large moving average… Show more
DEN-5887: Moved some load cell functions from dialysate outlet pump controller to DG Interface module.  Added small and large moving average filters to load cell readings.  Use small moving average for normal UF tracking/control and use large moving average for baseline and final reservoir weights to minimize reservoir transition errors due to load cell noise.