hdfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Stagi
DENBUG-328 updated the air trap event and the VectorCAST test cases

DENBUG-328 changed the get raw air trap function to get air trap function for alarm 215. Added air trap state change events

HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.c
    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.h
Merged DENBUG-328

RESOLVED IN CODEWALKTHROUGH

RESOLVED IN CODEWALKTHROUGH

DENBUG-328 addressed the code review comments

Done

Done

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

Please add the comments mentioning the reason/purpose for the checking fluid level timeout alarm?

DENBUG-328 updated the VectorCAST test cases

DENBUG-328 updated the UI institutional record structure

DENBUG-328 updated the min venous wide window range code

    • -4
    • +5
    /firmware/App/Modes/ModeTreatmentParams.c
    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.h
DENBUG-328 updated the air trap upper level fluid detection code

DENBUG-328 added persistent alarm for the air trap upper level fluid detection

    • -1
    • +23
    /firmware/App/Controllers/AirTrap.c
DENBUG-328 added min venous limit from the institutional record

    • -0
    • +16
    /firmware/App/Modes/ModeTreatmentParams.c
    • -0
    • +1
    /firmware/App/Modes/ModeTreatmentParams.h
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
HD-DENBUG-328_Venous Pressure Alarm Did Not Trigger When Pressure Dropped TO 0 Mmhg Staging
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

How is this if different than the else above? I don't think we need a separate if/else for not detected debounce - everything can be done in the if/else above.

How is this if different than the else above? I don't think we need a separate if/else for not detected debounce - everything can be done in the if/else above.

These should be initialized to zero in init function.

These should be initialized to zero in init function.