HD-DEN-11114_Dialysate Temperature Control (2 of 2)
HD-DEN-11114_Dialysate Temperature Control (2 of 2)
Should only set this to true when we know we will accept request.

Should only set this to true when we know we will accept request.

Comment should change from 200 to 50.

Comment should change from 200 to 50.

Done.

Done.

You can remove this line here. That pump has not yet been run.

You can remove this line here. That pump has not yet been run.

Fixed. Before transition from dialyzer state to wet selftests state, the VDO pump should be stopped.

Fixed. Before transition from dialyzer state to wet selftests state, the VDO pump should be stopped.

Move this line to when prime dialyzer state is completed before we go to wet self tests.

Move this line to when prime dialyzer state is completed before we go to wet self tests.

Done

Done

Fixed

Fixed

I believe reservoir 2 is used. static PRE_TREATMENT_RESERVOIR_MGMT_STATE_T handlePreTreatmentReservoirMgmtStartState( void ) { PRE_TREATMENT_RESERVOIR_MGMT_STATE_T state = PRE_TREATMENT_RESERVOIR_...

I believe reservoir 2 is used.

static PRE_TREATMENT_RESERVOIR_MGMT_STATE_T handlePreTreatmentReservoirMgmtStartState( void )
{
PRE_TREATMENT_RESERVOIR_MGMT_STATE_T state = PRE_TREATMENT_RESERVOIR_MGMT_START_STATE;

if ( TRUE == fillReservoirOneStartRequested )

Unknown macro: { fillReservoirOneStartRequested = FALSE; state = PRE_TREATMENT_RESERVOIR_MGMT_DRAIN_CMD_STATE; cmdSetDGActiveReservoir( DG_RESERVOIR_2 ); }


return state;
}

Fixed

Fixed

Done

Done

Done

Done

This might be reservoir 1 - there is a pre-treatment reservoir mgmt function handling this - check to see which reservoir is active at beginning of prime here.

This might be reservoir 1 - there is a pre-treatment reservoir mgmt function handling this - check to see which reservoir is active at beginning of prime here.

Why reset start time here?

Why reset start time here?

I think this needs to be closed or saline will not be forced across dialyzer into dialysate flow path.

I think this needs to be closed or saline will not be forced across dialyzer into dialysate flow path.

Make it clearer that this is for prime dialyzer state.

Make it clearer that this is for prime dialyzer state.

Dialyzer is spelled with a "z" not "s".

Dialyzer is spelled with a "z" not "s".

I would rename to something like prime dialyzer stop weight.

I would rename to something like prime dialyzer stop weight.

HD-DEN-11098_Implement HD Pre Treatment Mode Update
HD-DEN-11098_Implement HD Pre Treatment Mode Update
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH. (On behalf of Behrouz NematiPour who is on vacation)

RESOLVED in CODE WALKTHROUGH. (On behalf of Behrouz NematiPour who is on vacation)

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.