DEN-14197: Start DPo pump to same PWM duty cycle as DPi when in closed loop mode.

    • -1
    • +5
    /firmware/App/Controllers/DialOutFlow.c
If this condition (timeout) occurs in middle of pressure build up, will we release on transition to next state?

If this condition (timeout) occurs in middle of pressure build up, will we release on transition to next state?

Magic # 5 - use #define. What happens if prior bubble clear is still active when this times out? Seems like condition should check that bubbleClearActive is FALSE too.

Magic # 5 - use #define.
What happens if prior bubble clear is still active when this times out? Seems like condition should check that bubbleClearActive is FALSE too.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14307-dg_hd_dev-hd_dg_dvt-update-part-6' into develop

DEN-14307 updated the alarms name and the checkAlarms.sh to check the manual alarms table and fail upon compilation

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14099-bubbles-in-dialysate-out-line-from-dg-to-cartridge' into develop

DEN-14099 Add todo note for possible mitigation of valve closed too long.

DEN-14099 Fix pressure reading to use Venous and instant value.

HD-DEN-14197_SW S83 MG PI Optimization
HD-DEN-14197_SW S83 MG PI Optimization
DEN-14197: Range check DPi initial PWM duty cycle before setting.

    • -10
    • +7
    /firmware/App/Controllers/DialInFlow.c
HD-DEN-14099_Bubbles IN Dialysate Out Line From DG TO Cartridge
HD-DEN-14099_Bubbles IN Dialysate Out Line From DG TO Cartridge
HD-DEN-14307_DG HD Dev HD DG Dvt Update Part 6
HD-DEN-14307_DG HD Dev HD DG Dvt Update Part 6
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Controllers/DGInterface.c
    • -2
    • +2
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.h
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-14150

    • -2
    • +2
    /firmware/App/Controllers/DGInterface.c
    • -1
    • +1
    /firmware/App/Controllers/DGInterface.h
    • -16
    • +18
    /firmware/App/Controllers/DialInFlow.c
    • -2
    • +2
    /firmware/App/Controllers/SyringePump.c
    • -1
    • +1
    /firmware/App/Controllers/Temperatures.c
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14150 addressed a code review comment

DEN-14150 reversed the previous code changes. Calling resetBloodPumpRotorCount and resetDialInPumpRotorCount

Done

Done

DEN-14150 addressed a code review comment by making resetSignalFlags and resetAlarmSignalFlags functions to be called in pretreatment

    • -35
    • +30
    /firmware/App/Modes/ModeTreatment.c
DEN-14150 removed a blank line

Need to make sure these only get called once before pre-treatment.

Need to make sure these only get called once before pre-treatment.

DEN-14150 merged staging

    • -28
    • +131
    /firmware/App/Controllers/DGInterface.c
    • -2
    • +3
    /firmware/App/Controllers/DialInFlow.h
    • -30
    • +226
    /firmware/App/Modes/ModeStandby.c
    • -5
    • +21
    /firmware/App/Services/SystemComm.c
DEN-14099 Added pressure / release cycles to remove air bubbles during prime dialyzer.