DEN-14549: removed unneccessary pre-tx prime state.

DEN-14457-reservoir-drains-before-treatment-with-pre-treatment-build-flags-off

DEN-14457-reservoir-drains-before-treatment-with-pre-treatment-build-flags-off dragged into current branch

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

This config check is unnecessary and should be removed. The vars should always be initialized. Enable/Disable Alarms are done when checking the direction, etc.

This config check is unnecessary and should be removed. The vars should always be initialized. Enable/Disable Alarms are done when checking the direction, etc.

Merge branch 'staging' into DEN-14561-_release_-preprocessor-flag-specifies-code-that-conflicts-with-sw_config_-debug

HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug

Corrected _RELEASE_ logic, release build executes the same block as non debug selection

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c
    • -10
    • +11
    /firmware/App/Services/AlarmMgmt.c
They're kind of different. The stop function is fairly straightforward and is part of syringe pump driver. The start function is a lot more complex as it needs to figure out what it should be doing...

They're kind of different. The stop function is fairly straightforward and is part of syringe pump driver. The start function is a lot more complex as it needs to figure out what it should be doing and is part of treatment mode module.

Inconsistent naming here. StopSyringePump, StartHeparinPump. Suggest using same name for both.

Inconsistent naming here. StopSyringePump, StartHeparinPump. Suggest using same name for both.

Done.

Done.

DEN-14549: Addressed code review comments.

Done.

Done.

Done.

Done.

DEN-14549: maintain blood flow rate when alarm w/ blood recirc property pauses treatment. maintain Heparin dispense when blood recirc is occurring during alarm pause of treatment. include blood prime state in checks for inline pressure range.

    • -24
    • +4
    /firmware/App/Controllers/PresOccl.c
    • -18
    • +24
    /firmware/App/Modes/TreatmentStop.c
HD-DEN-14328_HD Event Issues
HD-DEN-14328_HD Event Issues
Add pressureSelfTestNormalizedStartTime to header.

Add pressureSelfTestNormalizedStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14549: Misc updates to dry self-tests in pre-treatment.

Merge branch 'DEN-14457-reservoir-drains-before-treatment-with-pre-treatment-build-flags-off' into DEN-14561-_release_-preprocessor-flag-specifies-code-that-conflicts-with-sw_config_-debug

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Reverted volume changes. Using a lower volume appeared to cause flowmeter test errors. Volumes match the original code, now.

HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug
HD-DEN-14561_Release Preprocessor Flag Specifies Code That Conflicts With SW Config Debug
DEN-14549: Fixed issue where door open alarm would have condition cleared repeatedly in standby start state. Made top alarm determination done from scratch each time.

DEN-14549: Removed obsolete internal ADC channel. Updates internal ADC channel transfer functions. Misc pre-tx state optimizations.

    • -11
    • +7
    /firmware/App/Modes/ConsumableSelfTest.c
    • -25
    • +16
    /firmware/App/Modes/ModePreTreat.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces