DEN-14557 reversed the blood leak's pass/fail logic in the zero sequence

HD-DEN-14557_Dvt 012 HD Valve Transition Time Out Valve Position Out OF Target During Cartridge
HD-DEN-14557_Dvt 012 HD Valve Transition Time Out Valve Position Out OF Target During Cartridge
DEN-14557 added code to the HD valves to accept a transition command if the valve is currently is not close to target

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

DEN-14011

Done.

Done.

Done.

Done.

Extra white space.

Extra white space.

Should align

Should align

DEN-14561 Wait for dry self test before final fill. Use reservoir 2 for Prime dialysate bypass.

If going back to start state, set pump track closed required, but not door closed. If skipping dry self-tests, set both to be required closed. Also, this s/w config looks like it's setup wrong. Fo...

If going back to start state, set pump track closed required, but not door closed. If skipping dry self-tests, set both to be required closed.

Also, this s/w config looks like it's setup wrong. For release build, we want to go back to start state.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

transition and resume part of stopped state should set pump track required so we don't have to keep setting it here every 50 ms.

transition and resume part of stopped state should set pump track required so we don't have to keep setting it here every 50 ms.

Should set door and pump track required here before we move on to next state.

Should set door and pump track required here before we move on to next state.

Should set pump track closed required, but not door here.

Should set pump track closed required, but not door here.

DEN-14011: implemented proper exit condition for dry self start state.

Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Do we need an else where we reset 10 second start time (if latch opened)?

Do we need an else where we reset 10 second start time (if latch opened)?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.