hdfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Fixed.

Fixed.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Changing from U32 to S32. It is a relative ms timer (so wouldn't be negative), but I needed to initialize this to something less than zero so that alarm ranking logic will be forced to rank a brand...

Changing from U32 to S32. It is a relative ms timer (so wouldn't be negative), but I needed to initialize this to something less than zero so that alarm ranking logic will be forced to rank a brand new alarm (triggered 0 ms ago).

What's the significance of changing to U32? Is this being used as a timer?

What's the significance of changing to U32? Is this being used as a timer?

Should venPres < venLowLimit be done outside if statement?

Should venPres < venLowLimit be done outside if statement?

HD-DEN-14557_Dvt 012 HD Valve Transition Time Out Valve Position Out OF Target During Cartridge
HD-DEN-14557_Dvt 012 HD Valve Transition Time Out Valve Position Out OF Target During Cartridge
HD-DEN-14999_Alarm Table Updates
HD-DEN-14999_Alarm Table Updates
Add timeWaitToFillMS to header.

Add timeWaitToFillMS to header.

HD-DEN-14788_DIAL_242 Air Bubble Sensor
HD-DEN-14788_DIAL_242 Air Bubble Sensor
HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
HD-DEN-14888_Unable TO Trigger Alarm 47 Ultrafiltration Rate IS Too High Error During Treatment2
I think we need to re-visit this. Could impact dialysate heating negatively.

I think we need to re-visit this. Could impact dialysate heating negatively.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.