Merge branch 'staging' into DEN-15049-fw-hd-voltage-not-trigger-alarm-76-when-voltage-out-of-range-in-line

I think we can just say "if ( TRUE == hasPowerBeenLost )" here. Zero trigger persistence time now that CPLD is making determination.

I think we can just say "if ( TRUE == hasPowerBeenLost )" here. Zero trigger persistence time now that CPLD is making determination.

I think we can fit this whole comment on the 1 line. Bring closer to break; if need more room.

I think we can fit this whole comment on the 1 line. Bring closer to break; if need more room.

No longer needed?

No longer needed?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Remove this method of determine whether AC power has been lost. CPLD function above is what we want.

Remove this method of determine whether AC power has been lost. CPLD function above is what we want.

Corrected

Corrected

We want to use CPLD for this. Keep this line, remove line below where initialized to FALSE.

We want to use CPLD for this. Keep this line, remove line below where initialized to FALSE.

DEN-15049 Comment correction / clarification

Should be dgPOSTPassed.

Should be dgPOSTPassed.

Update header.

Update header.

Still wrong though.

Still wrong though.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15105-dialyzer-flip' into develop

DEN-15105: rolled back req for DG POST to pass to start treatment workflow for now - not working.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15105-dialyzer-flip' into develop

    • -0
    • +6
    /firmware/App/Controllers/DGInterface.c
DEN-15105: Fix for DEN-15154 - saline bolus issue.

    • -2
    • +3
    /firmware/App/Modes/ModeTreatmentParams.c
DEN-15105: Resolved some compiler warnings. Require POST to pass before starting treatment workflow.

    • -0
    • +6
    /firmware/App/Controllers/DGInterface.c
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-15024 Use windowed timer for Syringe pump off alarm.

    • -8
    • +4
    /firmware/App/Controllers/SyringePump.c
HD-DEN-15024_FW Srshd 1333 Vtchd 283 Alarm ID HD Syringe Pump Not Stopped Error ID 111 Does Not
HD-DEN-15024_FW Srshd 1333 Vtchd 283 Alarm ID HD Syringe Pump Not Stopped Error ID 111 Does Not
Updated.

Updated.

DEN-15049 Correct header

Comment is identical to the DG implementation.

Comment is identical to the DG implementation.

DEN15049 Correct header

Merge branch 'staging' into DEN-15049-fw-hd-voltage-not-trigger-alarm-76-when-voltage-out-of-range-in-line

# Conflicts:

# firmware/App/Controllers/Voltages.c

    • -235
    • +244
    /firmware/App/Controllers/Voltages.c