I think we can just say "if ( TRUE == hasPowerBeenLost )" here. Zero trigger persistence time now that CPLD is making determination.

I think we can just say "if ( TRUE == hasPowerBeenLost )" here. Zero trigger persistence time now that CPLD is making determination.

I think we can fit this whole comment on the 1 line. Bring closer to break; if need more room.

I think we can fit this whole comment on the 1 line. Bring closer to break; if need more room.

No longer needed?

No longer needed?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Remove this method of determine whether AC power has been lost. CPLD function above is what we want.

Remove this method of determine whether AC power has been lost. CPLD function above is what we want.

Corrected

Corrected

We want to use CPLD for this. Keep this line, remove line below where initialized to FALSE.

We want to use CPLD for this. Keep this line, remove line below where initialized to FALSE.

Should be dgPOSTPassed.

Should be dgPOSTPassed.

Update header.

Update header.

Still wrong though.

Still wrong though.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

HD-DEN-15024_FW Srshd 1333 Vtchd 283 Alarm ID HD Syringe Pump Not Stopped Error ID 111 Does Not
HD-DEN-15024_FW Srshd 1333 Vtchd 283 Alarm ID HD Syringe Pump Not Stopped Error ID 111 Does Not
Updated.

Updated.

Comment is identical to the DG implementation.

Comment is identical to the DG implementation.

Comments appear to be swapped.

Comments appear to be swapped.

Update header

Update header

HD-DEN-14546_Add Option TO Resume Treatment Without Heparin For Empty Syringe Alarm
HD-DEN-14546_Add Option TO Resume Treatment Without Heparin For Empty Syringe Alarm
HD-DEN-15110_Always Use Max Rate OF 800Ml Min For Reservoir Fill
HD-DEN-15110_Always Use Max Rate OF 800Ml Min For Reservoir Fill
HD-DEN-15105_Dialyzer Flip
HD-DEN-15105_Dialyzer Flip
HD-DEN-15101_Alarm 139 HD AC Power Alarm Lost Triggers ON Startup
HD-DEN-15101_Alarm 139 HD AC Power Alarm Lost Triggers ON Startup
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH