DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adjusted Alarm limit

DEN-15189 Add activateSafetyShutdown to alarm ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE (62) if position C

Add treatmentTimeBroadcastTimerCtr to header.

Add treatmentTimeBroadcastTimerCtr to header.

DEN-15024 Change syringe pump off error count to max time windowed count.

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c
Add doorOpened to header.

Add doorOpened to header.

DEN-15183: indicate treatment over in treatment times broadcast if alarm preventing alarm resumption.

HD-DEN-15189_Safety Shutdown Not Activating When HD Valve Position Out OF Range Alarm 62 Occurs
HD-DEN-15189_Safety Shutdown Not Activating When HD Valve Position Out OF Range Alarm 62 Occurs
HD-DEN-15196_Alarm 98 Rework
HD-DEN-15196_Alarm 98 Rework
DEN-15183: Removed pressure limit alarm exemption for pump ramping. Address sequence issue with invert dialyzer logic DEN-15195.

    • -14
    • +0
    /firmware/App/Controllers/PresOccl.c
DEN-15104 Minor correction

DEN-14980-unable-to-resume-after-hd-unit-plugged-back-in-pre-treatment-after-priming

Adding an alarm blocking state to prevent cascade of errors during power fail.

    • -53
    • +75
    /firmware/App/Services/AlarmMgmt.c
HD-DEN-14980_Unable TO Resume After HD Unit Plugged Back IN Pre Treatment After Priming
HD-DEN-14980_Unable TO Resume After HD Unit Plugged Back IN Pre Treatment After Priming
HD-DEN-14325_FW Blood Leak Alarm Triggers When Cartridge Lever IS Unlocked
HD-DEN-14325_FW Blood Leak Alarm Triggers When Cartridge Lever IS Unlocked
DEN-15183: Addressed DEN-15191 by zeroing treatment time remaining in treatment times broadcast msg if an alarm has blocked resumption of treatment.

DEN-14183: Addressed DEN-15190 by allowing rinseback option regardless of whether alarm condition is cleared.

DEN-15024 Fix order of saving last syringe position and syringe detection.

    • -3
    • +6
    /firmware/App/Controllers/SyringePump.c
Merge branch 'staging' into DEN-15104-hd-fpga-communications

    • -91
    • +116
    /firmware/App/Services/FPGA.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-14325-fw-blood-leak-alarm-triggers-when-cartridge-lever-is-unlocked' into develop

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-15183 to develop.

    • -2
    • +2
    /firmware/App/Controllers/DGInterface.c
DEN-15183: Fix for DEN-14312 (DPi running when alarm in rinseback should stop DPi).

DEN-14325 added conditions to not check for blood leak alarm, air trap illegal level alarm, and the bubble detection alarm and self test request if the pump track is open

HD-DEN-15183_IN Treatment Rinseback Workflow Changes
HD-DEN-15183_IN Treatment Rinseback Workflow Changes
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Controllers/DGInterface.c
    • -2
    • +2
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.c
  1. … 7 more files in changeset.