Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Controllers/SyringePump.c
    • -2
    • +2
    /firmware/App/Controllers/SyringePump.h
    • -2
    • +2
    /firmware/App/Modes/ModeTreatmentParams.c
Merge branch 'DEN-15023-variation-in-syringe-pump-position-between-dvt-units' into staging

DEN-15023: fixes

    • -2
    • +2
    /firmware/App/Controllers/SyringePump.c
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-15203: code review comments.

    • -7
    • +12
    /firmware/App/Controllers/SyringePump.c
    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.c
Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Add blank line between ifs.

Add blank line between ifs.

Remove extra blank line.

Remove extra blank line.

Should this be looking at >2.5V (instead of pre-tx occlusionDetected flag)?

Should this be looking at >2.5V (instead of pre-tx occlusionDetected flag)?

This appears to be in ms. Should at least say in comment.

This appears to be in ms. Should at least say in comment.

Merge branch 'staging' into DEN-15023-variation-in-syringe-pump-position-between-dvt-units

HD-DEN-14262_Vba Homing Fails When Cartridge IS Loaded ON Device
HD-DEN-14262_Vba Homing Fails When Cartridge IS Loaded ON Device
Changed to 10 along with the max windowed timer count. DVT-005 was failing with a count of 5-7. It is passing now with max of 10.

Changed to 10 along with the max windowed timer count. DVT-005 was failing with a count of 5-7. It is passing now with max of 10.

The suggested mechanism to resume treatment without heparin had been to allow resume with minor flag changes to allow the treatment to continue. At this point clearing the alarm may be all that is ...

The suggested mechanism to resume treatment without heparin had been to allow resume with minor flag changes to allow the treatment to continue. At this point clearing the alarm may be all that is necessary. Code updated.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15024-fw-srshd-1333-vtchd-283-alarm_id_hd_syringe_pump_not_stopped_error-id-111-does-not' into develop

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c