Merged DEN-15396

DEN-15396 addressed a code review comment

    • -2
    • +3
    /firmware/App/Controllers/DialInFlow.c
    • -4
    • +1
    /firmware/App/Controllers/DialOutFlow.c
HD-DEN-15396_Alarm 42 HD Fault Dpo Rotor Encountered IN Blood Prime Dvt 001 2
HD-DEN-15396_Alarm 42 HD Fault Dpo Rotor Encountered IN Blood Prime Dvt 001 2
DEN-15396 opened up the persistent time of the DPo pumps rotor speed check

    • -1
    • +4
    /firmware/App/Controllers/DialOutFlow.c
    • -1
    • +2
    /firmware/App/Controllers/Temperatures.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15375

DEN-15375 fixed the implicit compile issues

Merged DEN-15375

DEN-15375 addresed the release fixes

HD-DEN-15375_FW Dialysate Heating Fixes
HD-DEN-15375_FW Dialysate Heating Fixes
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15363 Add windowed timer

    • -0
    • +12
    /firmware/App/Controllers/BloodLeak.c
HD-DEN-15250_Pre Treatment Does Not End Successfully
HD-DEN-15250_Pre Treatment Does Not End Successfully
This doesn't seem like a fix to root cause - may not work if user clears alarm quickly. Suggest looking at valve driver where alarm 62 is detected/triggered and maybe zeroing positionOutOfRangeCoun...

This doesn't seem like a fix to root cause - may not work if user clears alarm quickly.
Suggest looking at valve driver where alarm 62 is detected/triggered and maybe zeroing positionOutOfRangeCounter (reset 1 sec persistence) if AC power is out. That way, when AC is restored, valves will have a whole second to get where they're supposed to be before alarm is triggered.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15060-2 to staging.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15060-2-nfr-alarms-lower-priority-alarm-de-silence-higher-priority-alarm-audio' into develop

DEN-15060-2: Another fix to alarm silence functionality.

HD-DEN-15363_FW Handle Bld Sensor Communication Errors
HD-DEN-15363_FW Handle Bld Sensor Communication Errors
HD-DEN-15060_2 Nfr Alarms Lower Priority Alarm DE Silence Higher Priority Alarm Audio
HD-DEN-15060_2 Nfr Alarms Lower Priority Alarm DE Silence Higher Priority Alarm Audio
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15060-2-nfr-alarms-lower-priority-alarm-de-silence-higher-priority-alarm-audio' into develop