DEN-15359: Addressed DEN-15381.

    • -18
    • +29
    /firmware/App/Modes/ModeStandby.c
DEN-15359: Addressed DEN-15284.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Modes/ModeTreatmentParams.c
HD-DEN-15359_Usability Arterial And Venous Pressure Alarm Limits
HD-DEN-15359_Usability Arterial And Venous Pressure Alarm Limits
Merged DEN-15250 to staging.

Merge staging to DEN-15250.

DEN-15359: Updated tx params to change art/ven pressure limit params from min/max to window.

    • -27
    • +35
    /firmware/App/Controllers/PresOccl.c
    • -64
    • +17
    /firmware/App/Modes/ModeTreatment.c
    • -44
    • +14
    /firmware/App/Modes/ModeTreatmentParams.c
    • -8
    • +4
    /firmware/App/Modes/ModeTreatmentParams.h
The valves are not returned to their target position until the alarm is cleared. I could make a broader change, but is that what you're suggesting?

The valves are not returned to their target position until the alarm is cleared. I could make a broader change, but is that what you're suggesting?

DEN-13865: Addressed code review comment.

    • -1
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-15396 Use greater of 10sec or 10 Rotor turn times for minimum persistent error time.

    • -4
    • +8
    /firmware/App/Controllers/DialInFlow.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-13865-fw-duplicate-alarms-occur-during-post-treatment-rinseback-and-block-workflow' into develop

DEN-13865: More fixes to alarm re-trigger prevention logic.

Merge branch 'DEN-13865-fw-duplicate-alarms-occur-during-post-treatment-rinseback-and-block-workflow' into develop

DEN-13865: fix re-trigger block logic.

    • -20
    • +23
    /firmware/App/Services/AlarmMgmt.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-13865-fw-duplicate-alarms-occur-during-post-treatment-rinseback-and-block-workflow' into develop

DEN-13865: Handle no re-trigger action or end-tx only alarms during rinseback/recirc states of treatment mode.

Merge branch 'DEN-13865-fw-duplicate-alarms-occur-during-post-treatment-rinseback-and-block-workflow' into develop

    • -0
    • +1
    /firmware/App/Controllers/SyringePump.c
DEN-13865: Implemented alarm system feature to prevent alarm re-trigger per new properties. Addressed DEN-15295.

    • -0
    • +1
    /firmware/App/Controllers/SyringePump.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

DEN-15367 updated the use wet cartrdige test configuration

So then can we remove power loss alarm active checks from this condition (just keep the cpld power loss check as before)?

So then can we remove power loss alarm active checks from this condition (just keep the cpld power loss check as before)?

Do we need the alarm active checks? Isn't power loss check sufficient?

Do we need the alarm active checks? Isn't power loss check sufficient?

HD-DEN-13865_FW Duplicate Alarms Occur During Post Treatment Rinseback And Block Workflow
HD-DEN-13865_FW Duplicate Alarms Occur During Post Treatment Rinseback And Block Workflow
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15250-pre-treatment-does-not-end-successfully' into develop