DEN-15396 Code review comment, use macro.

    • -4
    • +2
    /firmware/App/Controllers/DialInFlow.c
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Controllers/DialInFlow.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

DEN-15306-hd-fault-valve-position-alarm-after-restoring-ac-power

Aligned tests.

Merge branch 'DEN-15396-2-alarm-42-hd-fault-dpo-rotor-encountered-in-blood-prime-dvt-001' into staging

Merge branch 'DEN-15396-2-alarm-42-hd-fault-dpo-rotor-encountered-in-blood-prime-dvt-001' into develop

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15306-hd-fault-valve-position-alarm-after-restoring-ac-power

Made the power loss check uniform

isACPowerLoss is an alarm status, move to AlarmMgmt

Trigger with CPLD power loss flag, recover HW state when AC returns, start alarm block countdown after alarm is cleared.

    • -2
    • +2
    /firmware/App/Controllers/DGInterface.c
    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -18
    • +25
    /firmware/App/Controllers/Valves.c
Add stabilizationStartTimeMs to header.

Add stabilizationStartTimeMs to header.

DEN-15359: handle blood flow rate change in middle of stabilization state.

    • -2
    • +10
    /firmware/App/Controllers/PresOccl.c
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Can't keep in message x47 as f/w will not have calculated new min/max yet. I made it so message x9 will broadcast immediately after user changes pressure limit windows.

Can't keep in message x47 as f/w will not have calculated new min/max yet.
I made it so message x9 will broadcast immediately after user changes pressure limit windows.

DEN-15359: broadcast pressure data immediately after window change by user.

Seems the only message which is providing the calculated A/V min/max is here in message 0x09. I think it was better to keep the min, max of the arterial and venous in the message 0x47 (A/V InTx adj...

Seems the only message which is providing the calculated A/V min/max is here in message 0x09.
I think it was better to keep the min, max of the arterial and venous in the message 0x47 (A/V InTx adj) to let UI immediately have the updated values,
instead of the x09 which is a broadcast and won't immediately send the updated/calculated min/max, and causes a short delay (depending on interval of the broadcast) to update the UI.
Or in both if the message 0x09 is getting continuously updated during treatment.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Fixed.

Fixed.

DEN-15359: Addressed code review comments.

    • -0
    • +3
    /firmware/App/Modes/ModeTreatmentParams.c
Fixed.

Fixed.

Yes. Fixed.

Yes. Fixed.

Yes. Fixed.

Yes. Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

same here?

same here?