RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

We should send this countdown msg from a publish function that gets called by exec function for treatment end only once per second (see how this msg is sent by rinseback mode for example). And shou...

We should send this countdown msg from a publish function that gets called by exec function for treatment end only once per second (see how this msg is sent by rinseback mode for example).
And should send the message all the time. If in the wait for rinseback state, just send zeroes (and UI won't show countdown). If in paused state, always send 300 seconds for timeout and 300-(bloodSittingTimerCtr / 20) for countdown (don't let it go negative or wrap though). If an alarm dialog is being displayed, it should show the countdown for the whole 5 minutes (not just the last minute after warning).

HD-DEN-15664_Blood Pump Stop Time Exceeded Alarms 124 125 165 And 212 Alarm Timer Feature IS
HD-DEN-15664_Blood Pump Stop Time Exceeded Alarms 124 125 165 And 212 Alarm Timer Feature IS
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Add resumeRinsebackRequested.

Add resumeRinsebackRequested.

HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done

Done

Done.

Done.

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Blood pump would be stopped in this state, correct? So we need to call handleTreatmentStopBloodSittingTimer().

Done

Done

Move up with other setup functions.

Move up with other setup functions.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Is this being permanently removed. I believe there is a requirement for generating this alarm.

Should be the recovery alarm.

Should be the recovery alarm.