HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus
HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus
HD-DEN-15893_Cartridge Lever IS Open Alarm When Cartridge Lever IS Closed
HD-DEN-15893_Cartridge Lever IS Open Alarm When Cartridge Lever IS Closed
HD-DEN-15719_Blood Pump Stopped Alarms 211 212 124 125 Not Relevant For Alarms That Force TX TO
HD-DEN-15719_Blood Pump Stopped Alarms 211 212 124 125 Not Relevant For Alarms That Force TX TO
HD-DEN-15804_Blood Flow Rate Accuracy Results Fail OR Are Close TO Failing
HD-DEN-15804_Blood Flow Rate Accuracy Results Fail OR Are Close TO Failing
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Fixed. Thank you!

Fixed. Thank you!

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

The trimmer heater is turned back on in the transition back to dialysis: transitionToDialysis( ) This is definitely a place I would suggest refactoring, as the trimmer heater should follow dialysa...

The trimmer heater is turned back on in the transition back to dialysis: transitionToDialysis( )

This is definitely a place I would suggest refactoring, as the trimmer heater should follow dialysate flow everywhere, but is somewhat distributed.

Where is the heater turned back on? Is the state saved?

Where is the heater turned back on? Is the state saved?

Is this intended to be a BITWISE OR? Did you intend || ?

Is this intended to be a BITWISE OR? Did you intend || ?

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed overrides.

Removed overrides.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CDOE WALKTHROUGH

RESOLVED IN CDOE WALKTHROUGH

Done

Done

Done

Done

Add endTreatmentRequested

Add endTreatmentRequested