hdfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Resume requested flag gets set elsewhere (from user clearing alarm -> CAN -> AlarmMgmt -> Modes -> this module) and should not be set here.

Resume requested flag gets set elsewhere (from user clearing alarm -> CAN -> AlarmMgmt -> Modes -> this module) and should not be set here.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

So this is only going to stop pumps/heater if both block recircs are active. I don't think that's what you want. My understanding of what you're trying to do here is this: 1) if dialysate recirc is...

So this is only going to stop pumps/heater if both block recircs are active. I don't think that's what you want.
My understanding of what you're trying to do here is this:
1) if dialysate recirc is blocked by an active alarm (includes user stop alarm), we should stop DPi and HR. Otherwise, we should run DPi and HR.
2) if user stop alarm is active or we are in any state other than recirc state, we should stop BP. Otherwise, we should run BP to recirculate blood side of circuit.

I don't think you need to repeat @details Inputs: and @details Outputs:. Only need those for first row. See other functions for example.

I don't think you need to repeat @details Inputs: and @details Outputs:. Only need those for first row. See other functions for example.

Blank line before banner comment.

Blank line before banner comment.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Still more here than should be. Trimmer heater is being turned off twice now. DPo should already be stopped. And VBA/VBV should be handled by state machine I would think unless user stop requiremen...

Still more here than should be. Trimmer heater is being turned off twice now. DPo should already be stopped. And VBA/VBV should be handled by state machine I would think unless user stop requirement talks about valves too - and even if so, valves should only close when user stop alarm is active - nothing to do with recirc properties in alarms.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Change postponed for further testing. No change in BloodPumps. Done.

Change postponed for further testing. No change in BloodPumps.
Done.

Change postponed for further testing. No change in DialIn. Done.

Change postponed for further testing. No change in DialIn.
Done.

Change postponed for further testing. No change in DialOut. DONE.

Change postponed for further testing. No change in DialOut.
DONE.

Done.

Done.

Done.

Done.

BP is turned on in the resume from stop, and patient reconnect states, .100ml/min. Removed anything that is off, and stays off in recirculation. Done.

BP is turned on in the resume from stop, and patient reconnect states, .100ml/min. Removed anything that is off, and stays off in recirculation.
Done.

s/b recirculate. Done.

s/b recirculate. Done.

Done

Done

Done.

Done.

Agree that BP is on in the recirc state. Forget 2nd comment - focus on multiple function calls comment.

Agree that BP is on in the recirc state. Forget 2nd comment - focus on multiple function calls comment.

In treatment recirc, BP is turned on in the resume from stop, and patient reconnect states, .100ml/min. It seems as though the user stop should turn off both pumps.

In treatment recirc, BP is turned on in the resume from stop, and patient reconnect states, .100ml/min. It seems as though the user stop should turn off both pumps.

HD-DEN-15927_Alarm 127 Low Venous Pressure Triggers Post Treatment
HD-DEN-15927_Alarm 127 Low Venous Pressure Triggers Post Treatment