HD-DEN-16175_Issues Under Vtchd 635 TX Timer
HD-DEN-16175_Issues Under Vtchd 635 TX Timer
HD-DEN-16018_FW UF Control During Blood Leak Alarm Recover
HD-DEN-16018_FW UF Control During Blood Leak Alarm Recover
Remove commented / unused code. If it is to be used, it does not check for overflow.

Remove commented / unused code. If it is to be used, it does not check for overflow.

HD-DEN-15980_Pressure Alarm Windows Widened For Extended Period OF Time Due TO Frequency OF Air
HD-DEN-15980_Pressure Alarm Windows Widened For Extended Period OF Time Due TO Frequency OF Air
Add currentPostTreatmentState to header

Add currentPostTreatmentState to header

HD-DEN-14815_Cartridge Removal Screen IS Displayed Prematurely
HD-DEN-14815_Cartridge Removal Screen IS Displayed Prematurely
DGFault instead of Dfault

DGFault instead of Dfault

HD-DEN-16166_Event Message For Pass Fail Bld Self Test
HD-DEN-16166_Event Message For Pass Fail Bld Self Test
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Updated.

Updated.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Need to do same for pump track latch.

Need to do same for pump track latch.

Remove this comment.

Remove this comment.

HD-DEN-15993_During Treatment Alarms 167 And Door And Latch Open Offer Resume Option That
HD-DEN-15993_During Treatment Alarms 167 And Door And Latch Open Offer Resume Option That
DEN-16102: changed dry self start state to always handle alarms with stop property.
DEN-16102: changed dry self start state to always handle alarms with stop property.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Wouldn't we want to stop if an alarm occurred during recirculation, which affected the blood circuit? My inclination would be to manage which alarms were enabled in this state rather than not imple...

Wouldn't we want to stop if an alarm occurred during recirculation, which affected the blood circuit? My inclination would be to manage which alarms were enabled in this state rather than not implement the mechanism to stop the pumps.

Implementing the change.
Done.

Should blood recirc blocked be based on alarm properties (as you've written here) or based on whether user stopped (specifically)? My feeling is that since there is no blood in blood side of circui...

Should blood recirc blocked be based on alarm properties (as you've written here) or based on whether user stopped (specifically)? My feeling is that since there is no blood in blood side of circuit (because rinseback was completed before we got to recirc), the NoBloodRecirc alarm property maybe doesn't apply here and we should only stop BP if the user stop alarm is active.