HD-DENBUG-101_Dpi Does Not Run When Blood Leak Sensor Fails TO Zero
HD-DENBUG-101_Dpi Does Not Run When Blood Leak Sensor Fails TO Zero
I think we need the minimum check too. uFRate can be negative if user sets volume to something lower than what has already been collected (negative volume) which would yield a negative rate which i...

I think we need the minimum check too. uFRate can be negative if user sets volume to something lower than what has already been collected (negative volume) which would yield a negative rate which is < 0.0.

Check casting. Result is U32 but could be assigned an S32.

Check casting. Result is U32 but could be assigned an S32.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Fixed.

Fixed.

Fixed.

Fixed.

Add setUFRate to header

Add setUFRate to header

Add setUFRate to header.

Add setUFRate to header.

HD-DENBUG-96_Treatment Log Inaccuracies
HD-DENBUG-96_Treatment Log Inaccuracies
HD-DEN-16487_Dialin Config Helper Commands Are Not Supported With The Latest HD And DG Builds
HD-DEN-16487_Dialin Config Helper Commands Are Not Supported With The Latest HD And DG Builds
HD-DEN-16446_HD Dev Test Clean UP Vectorcast
HD-DEN-16446_HD Dev Test Clean UP Vectorcast
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Thanks. Was not sure about that but wanted to cover it if bolus extended to treatment mode. Removed.

Thanks. Was not sure about that but wanted to cover it if bolus extended to treatment mode. Removed.