hdfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DENBUG-298 addressed the saline bolus rejection conditions

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added code in saline bolus to reject saline bolus request if the active alarm has blood and dialysate recirculation blocked

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 fixed the 4th level for saline bolus in treatment stop

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 updated the code

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 addressed the code review comments

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added broadcast to saline bolus in treatment stop. Made the stop bolus available in treatment stop too. Made VBV to open in saline bolus

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 changed the location of the saline bolus request in treatment stop

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added code in saline bolus to hold the start flag to true in treatment stop so the treatment stop does not transition

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added code to accept saline bolus request in tretment stop as well

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added the treatment mode to the minimize code in alarm management

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo