Merge branch 'DEN-15196-DIAL-262-fixes' into develop

DIAL-262: fixed operation status event to have more accurate transitioning in between dialysis sub modes.

DEN-15558: Reversed temporary change in dialyzer prime BP speed. Increased max venous pressure from 300 to 400 mmHg.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Fixed merge conflict.

DEN-15558: temporary slow down of BP during pre-tx dialyzer prime state to see if helps with blood leak issue.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merged DEN-15459

    • -2
    • +2
    /firmware/App/Controllers/DialInFlow.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15459

DEN-15459 changed the blood leak alarm detection to not trigger if the HD mode is not in treatment

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -2
    • +2
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.h
    • -2
    • +2
    /firmware/App/Controllers/DialOutFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.h
Merged DEN-15278 to staging.

Merged staging to DEN-15278.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Modes/PreTreatmentRecirc.c
Merge branch 'DEN-15196-DIAL-262' into staging

DEN-15493 Update Rinseback flow rate range to 100-300, default 200. Old was 50-150, default 75.

    • -1
    • +1
    /firmware/App/Modes/ModeTreatmentParams.c
DEN-15278: Addressed code review comment.

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15278-blood-stop-air-trap-alarm-declares-during-saline-bolus-treatment' into develop

Merge branch 'staging' into DEN-15363-fw-handle-bld-sensor-communication-errors

DEN-15278: no countdown timer if alarm 165 (blood sitting too long) is active.

DEN-15367 addressed the code review comments

    • -0
    • +2
    /firmware/App/Controllers/DGInterface.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

    • -78
    • +152
    /firmware/App/Modes/ModeStandby.c
DEN-15440 added code to partial drain the reservoirs in patient connect state machine in pretreatment

DEN-14004 added another message to request and DG service record from HD that is different from UI