Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-302 added more comments

DENBUG-302 changed the UF rate that is reported to UI in mL/min

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-302 updated the code to check for the UF rate confirmation in operation mode. DEN-17237 changed the code to re-trigger the placeholder cleaning mode alarms if they got cleared

    • -22
    • +29
    /firmware/App/Modes/ModeTreatment.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

    • -1
    • +10
    /firmware/App/Modes/TreatmentRecirc.c
DENBUG-302 continued with the changes for the UF rate screen support

    • -55
    • +51
    /firmware/App/Modes/ModeTreatment.c
    • -15
    • +13
    /firmware/App/Modes/OperationModes.c
    • -1
    • +15
    /firmware/App/Modes/OperationModes.h
    • -1
    • +10
    /firmware/App/Modes/TreatmentRecirc.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-302 added code to ignore sending the UF rate to the user if the UF volume is 0

    • -14
    • +22
    /firmware/App/Modes/ModeTreatment.c
DENBUG-298 addressed the code review comment

DENBUG-302 updated the code to send the UF rate back to the user and wait for the user to accept or reject it

DENBUG-298 removed code to not countdown when in saline bolus and in treatment stop

DENBUG-302 changed the code to send the new calculated UF rate to the user to accpet or reject and then based on that the treatment duration is updated

    • -11
    • +88
    /firmware/App/Modes/ModeTreatment.c
    • -11
    • +40
    /firmware/App/Modes/OperationModes.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 changed the block saline bolus to rinse back from dialysate recirculation

DENBUG-298 addressed the code review comment

    • -11
    • +40
    /firmware/App/Modes/OperationModes.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 addressed the saline bolus rejection conditions

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 added code in saline bolus to reject saline bolus request if the active alarm has blood and dialysate recirculation blocked

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DENBUG-298-user-unable-to-deliver-saline-bolus-during-alarm-conditions-where-the-blood-pump-is-recirculating' into develop

DENBUG-298 fixed the 4th level for saline bolus in treatment stop