Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-16773 added the changes from staging

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-16773 removed unused code in the airpump.c

DEN-16773 addressed a code review comment

DEN-16773 removed alarm 318

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 116 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
DEN-15886: Added events for air trap fill and air pump on/off events.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 21 more files in changeset.
DEN-15105: Fix for DEN-15116.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 20 more files in changeset.
DEN-14999: Removed alarm 306. Increased t/o on alarm 318 to 10 sec. Change alarm 141 to not clear immediate, but instead clear when low level sees fluid.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 46 more files in changeset.
DEN-14549: Update to arterial/venous pressure limit alarm logic. Update measured UF volume in all dialysis states. Remove dialysis and UF start states.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 10 more files in changeset.
DEN-14344: removed venous high limit increase while air pump is operating. Increased air pump operation timeout to 6s to handle worst case bp ramping.

  1. … 1 more file in changeset.
DEN-14344: code review fixes.

  1. … 2 more files in changeset.
DEN-14344: code review fixes.

  1. … 2 more files in changeset.
DEN-14344: Misc formatting fixes.

  1. … 1 more file in changeset.
DEN-14344: complier fixes and warnings

  1. … 1 more file in changeset.
DEN-14344: created air pump monitor function for alarm id 316. Created a timer for venous pressure to handle adjusted treatment values for a time while the air pump is operating. Allow HD to start filling after consumable testing in pretreatment instead of waiting after dry self tests. formatted ifndef release code blocks to look neater.

  1. … 3 more files in changeset.
DEN-14344: implmented venous pressure limits to increase when air pump is on. Reworked Air trap logic to turn off after air in the upper sensor for 2 seconds. Changed air pump timeout to be 5s.

  1. … 2 more files in changeset.
DEN-14344: implementation with of alarm id 316

Auto stash before merge of "develop" and "origin/develop"

DEN-14344: Air pump integration into air trap and dry self tests.

  1. … 2 more files in changeset.
DEN-14344: Air Pump integration into air trap control

  1. … 1 more file in changeset.