DialOutFlow.h

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-5887: Moved some load cell functions from dialysate outlet pump controller to DG Interface module. Added small and large moving average filters to load cell readings. Use small moving average for normal UF tracking/control and use large moving average for baseline and final reservoir weights to minimize reservoir transition errors due to load cell noise.

  1. … 6 more files in changeset.
DEN-4641: Merged with master. Fixed merge conflicts. Addressed code review feedback.

  1. … 25 more files in changeset.
DEN-4641: Distributed remaining payload record definitions to associated modules and deleted MessagePayloads.h. Added messaging and verification of in-line pressure limit changes.

  1. … 14 more files in changeset.
DEN-4641: Changed HD pump Dialin set point override commands to simply set commands.

  1. … 6 more files in changeset.
DEN-4640: Addressed code review feedback.

DEN-4640: Further developed saline bolus functionality.

  1. … 12 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 60 more files in changeset.
DEN-4211: Finished HD pumps open loop Dialin command update. Started data infrustructure for treatment parameters mode.

  1. … 15 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 70 more files in changeset.
Added rotor speed to dialysate pumps. Updated doxygen commenting. Added CPP check settings for the HD firmware project.

  1. … 7 more files in changeset.
Further implementation of UF settings change messaging and handling. Code review fixes.

  1. … 17 more files in changeset.
Added HD specific common header. Streamlined includes for Common.h and HDCommon.h.

  1. … 55 more files in changeset.
Updated more doxygen comments.

  1. … 9 more files in changeset.
Tuned UF control. Updated doxygen comments for DialOutFlow and Dialysis modules.

  1. … 8 more files in changeset.
Added DPo implementation. Added some messagine for DPo. Added FIRFilters. Updated HALCoGen pinouts.

  1. … 38 more files in changeset.
Testing changes as per code review

  1. … 3 more files in changeset.
Checking for settings in parameters passed

  1. … 3 more files in changeset.
Added load cell simulation

  1. … 5 more files in changeset.
Update CAN interface to test DialOutFlow module

  1. … 8 more files in changeset.
Added DialOutFlow.c and h. It compiles, it is not functional

  1. … 1 more file in changeset.