PresOccl.c

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-14549: maintain blood flow rate when alarm w/ blood recirc property pauses treatment. maintain Heparin dispense when blood recirc is occurring during alarm pause of treatment. include blood prime state in checks for inline pressure range.

  1. … 3 more files in changeset.
Add pressureSelfTestNormalizedStartTime to header.

Add pressureSelfTestNormalizedStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14549: Misc updates to dry self-tests in pre-treatment.

  1. … 3 more files in changeset.
It is in the header under Outputs.

It is in the header under Outputs.

Add dgServiceAndUsageData to header.

Add dgServiceAndUsageData to header.

HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
HD-DEN-14549_System Fault After Resuming Heparin Then Starting Saline Bolus During Treatment
Fixed merge conflicts.

  1. … 17 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 10 more files in changeset.
Merge branch 'staging' into DEN-14344-sw-s85-mg-air-pump

  1. … 8 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 112 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

Fixed. Thanks!

For a release build, looks like we would do nothing and get stuck in this state.

For a release build, looks like we would do nothing and get stuck in this state.

Add doxygen comment for persistence. Try to capture thinking on why this is such a brief period.

Add doxygen comment for persistence. Try to capture thinking on why this is such a brief period.

Merge branch 'DEN-14344-sw-s85-mg-air-pump' into develop

  1. … 1 more file in changeset.
DEN-14344: removed venous high limit increase while air pump is operating. Increased air pump operation timeout to 6s to handle worst case bp ramping.

  1. … 1 more file in changeset.
Fixed. Thanks!

Fixed. Thanks!

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

The only time we would be adding more states is if we are no longer on HIGH/LOW control, like switching to PWM. That would be such a significant change that this would be reviewed.

The only time we would be adding more states is if we are no longer on HIGH/LOW control, like switching to PWM. That would be such a significant change that this would be reviewed.

Would rather you set a BOOL based on state and then pass the BOOL to persistent alarm function. I'm afraid we could add more states later and forget about this.

Would rather you set a BOOL based on state and then pass the BOOL to persistent alarm function. I'm afraid we could add more states later and forget about this.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.