DEN-14011: created handleDrySelfTestStartState. code review comments.

  1. … 1 more file in changeset.
Should this be moved into "if" body? Doesn't appear to need setting until we resume.

Should this be moved into "if" body? Doesn't appear to need setting until we resume.

Should this 10 seconds only start once latch is locked?

Should this 10 seconds only start once latch is locked?

Not sure exception for treatment stop state still makes sense. We can be running BP in that state now depending on alarm. Should still alarm on occlusion.

Not sure exception for treatment stop state still makes sense. We can be running BP in that state now depending on alarm. Should still alarm on occlusion.

HD-DEN-14011_Occulsion Alarm When Reseating
HD-DEN-14011_Occulsion Alarm When Reseating
DEN-14011: added condition to check if BP is running while we do occlusion checks.

DEN-14549: Rolled back change to include blood prime for normal inline pressure range checks. Leave as fixed max only range check in blood prime.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14549: include venous check in including blood prime.

Yes, may need to be done when merging to Staging. Is this flag used independently anywhere? Or planned to be?

Yes, may need to be done when merging to Staging. Is this flag used independently anywhere? Or planned to be?

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

I don't think I have this bug fix in my branch. Needs to be addressed in a more recent branch.

I don't think I have this bug fix in my branch. Needs to be addressed in a more recent branch.

Good catch. Updated Voltages.c to behave same as Battery.c (trigger the correction alarm depending on whether in a treatment mode).

Good catch. Updated Voltages.c to behave same as Battery.c (trigger the correction alarm depending on whether in a treatment mode).

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

This check for _IN_TREATMENT may need to be added to Valves.c ALARM_ID_HD_AC_POWER_LOST check from recent bug fix. Both look to be checked in every case. Is this flag used independently anywhere?

This check for _IN_TREATMENT may need to be added to Valves.c ALARM_ID_HD_AC_POWER_LOST check from recent bug fix. Both look to be checked in every case. Is this flag used independently anywhere?

Treatment mode check is not done in Voltages.c, so ALARM_ID_HD_AC_POWER_LOST will be set there in addition to ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT here. Is this the expected behavior?

Treatment mode check is not done in Voltages.c, so ALARM_ID_HD_AC_POWER_LOST will be set there in addition to ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT here. Is this the expected behavior?

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

They're kind of different. The stop function is fairly straightforward and is part of syringe pump driver. The start function is a lot more complex as it needs to figure out what it should be doing...

They're kind of different. The stop function is fairly straightforward and is part of syringe pump driver. The start function is a lot more complex as it needs to figure out what it should be doing and is part of treatment mode module.

Inconsistent naming here. StopSyringePump, StartHeparinPump. Suggest using same name for both.

Inconsistent naming here. StopSyringePump, StartHeparinPump. Suggest using same name for both.

Done.

Done.

Done.

Done.

Done.

Done.

DEN-14549: maintain blood flow rate when alarm w/ blood recirc property pauses treatment. maintain Heparin dispense when blood recirc is occurring during alarm pause of treatment. include blood prime state in checks for inline pressure range.

  1. … 3 more files in changeset.
Add pressureSelfTestNormalizedStartTime to header.

Add pressureSelfTestNormalizedStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

Add pressureSelfTestPreNormalStartTime to header.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

DEN-14549: Misc updates to dry self-tests in pre-treatment.

  1. … 3 more files in changeset.