PresOccl.c

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Fixed.

Fixed.

Variable is also updated by receipts of DG flow broadcasts. I believe it is overridable on DG side.

Variable is also updated by receipts of DG flow broadcasts. I believe it is overridable on DG side.

Concern was originally just for retract due to highest speed. We ended up reducing retract speed to reduce likelihood of stalls. Still a good idea to check for it though and since retract, preload ...

Concern was originally just for retract due to highest speed. We ended up reducing retract speed to reduce likelihood of stalls. Still a good idea to check for it though and since retract, preload and seek all share same relatively high speed, these are the ops that have the check. For bolus and continuous, should not stall at these speeds and would catch with rate alarm if it did.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

Fixed.

Shouldnt this be set to 0U instead of 0.0F?

Shouldnt this be set to 0U instead of 0.0F?

This should be U32.

This should be U32.

This should be U32.

This should be U32.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

RESOLVED IN CODE WALKTHROUGH.

Should we check for stall in bolus & continuous states?

Should we check for stall in bolus & continuous states?

seems odd to create a variable just for it to be returned. Is this value supposed to be overridable?

seems odd to create a variable just for it to be returned. Is this value supposed to be overridable?

Shouldnt this be set to 0U instead of 0.0F?

Shouldnt this be set to 0U instead of 0.0F?

HD-DEN-14906_Dialyzer DE Prime
HD-DEN-14906_Dialyzer DE Prime
Fixed.

Fixed.

Fixed.

Fixed.

This will hard fault the HD. Please change to getTreatmentParameterU32()

This will hard fault the HD. Please change to getTreatmentParameterU32()

This will hard fault the HD. Please change to getTreatmentParameterU32()

This will hard fault the HD. Please change to getTreatmentParameterU32()

HD-DEN-13154_Heparin Syringe Empty Position Incorrect
HD-DEN-13154_Heparin Syringe Empty Position Incorrect
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH