Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-12441 Merge staging

  1. … 1 more file in changeset.
Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

  1. … 2 more files in changeset.
DEN-12224 updated the blood leak with an exit normal state command

  1. … 1 more file in changeset.
One letter enums could be more descriptive with their names.

One letter enums could be more descriptive with their names.

Why no SW Config in Release?

Why no SW Config in Release?

Merged DEN-12224

  1. … 2 more files in changeset.
DEN-12224 updated the blood leak emebedded mode

  1. … 3 more files in changeset.
Done

Done

Done

Done

Done

Done

Done

Done

Yes. setupForBloodRecirculationState() does this on Init and entry to Blood Recirculation.

Yes. setupForBloodRecirculationState() does this on Init and entry to Blood Recirculation.

This was a place holder for NVRAM saving of Treatment Time. I removed it.

This was a place holder for NVRAM saving of Treatment Time. I removed it.

Is this timer reset to 0 when we go to states where blood is recirculating? I think it should be.

Is this timer reset to 0 when we go to states where blood is recirculating? I think it should be.

Put spaces between parenthesis and inner condition code. Prefer literals be on right side of compare operators except for "==".

Put spaces between parenthesis and inner condition code. Prefer literals be on right side of compare operators except for "==".

Consider reading these blocked flags into local booleans so you don't have to keep calling these functions (and avoid possibility that function returns different value in subsequent call).

Consider reading these blocked flags into local booleans so you don't have to keep calling these functions (and avoid possibility that function returns different value in subsequent call).

I think syringe pump is stopped in every scenario so maybe this should stay here instead of in various setup functions.

I think syringe pump is stopped in every scenario so maybe this should stay here instead of in various setup functions.

I think VDi and VDo valves are closed in every scenario so maybe they should just be set here instead of in the various setup functions.

I think VDi and VDo valves are closed in every scenario so maybe they should just be set here instead of in the various setup functions.

DEN-12224 updated the blood leak embedded mode

  1. … 3 more files in changeset.