DEN-14906: Increased HD pump max current persistence from 2 to 5 seconds. Added 1 minute delay following start/resume of dialysis where venous pressure limits are set to max selectable to allow pumps to stabilize.

  1. … 2 more files in changeset.
HD-DEN-14557_Dvt 012 HD Valve Transition Time Out Valve Position Out OF Target During Cartridge
HD-DEN-14557_Dvt 012 HD Valve Transition Time Out Valve Position Out OF Target During Cartridge
DEN-14557 updated the code set the valves status to position unknown if the tranistion to a position timedout.

DEN-14906: Fix response to Dialin requiring ACK when it should not.

  1. … 1 more file in changeset.
Merge branch 'staging' into DEN-14830-incorrect-dialysate-temperature-alarm-triggered

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-14788: code review comments.

DEN-14999: Reduced BP occlusion clear threshold from 5500 to 5000 above baseline.

HD-DEN-14999_Alarm Table Updates
HD-DEN-14999_Alarm Table Updates
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
HD-DEN-14788_DIAL_242 Air Bubble Sensor
HD-DEN-14788_DIAL_242 Air Bubble Sensor
DEN-14788: DIAL-242: added check for bubble sensor override.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 12 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Done

Done