Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-14906: fixed merge conflicts from staging.

  1. … 8 more files in changeset.
DEN-14999: Fixed issue with repeated trimmer heater commands being sent to DG.

  1. … 1 more file in changeset.
Merge branch 'DEN-14906-dialyzer-de-prime' into develop

  1. … 1 more file in changeset.
DEN-14906: Fixed arterial high pressure check while in adjunct treatment states.

DEN-14999: Dont trigger alarm 312 unless in treatment mode or pre-treatment mode after priming complete.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Fixed.

Fixed.

DEN-14906: Addressed code review comment.

Changing from U32 to S32. It is a relative ms timer (so wouldn't be negative), but I needed to initialize this to something less than zero so that alarm ranking logic will be forced to rank a brand...

Changing from U32 to S32. It is a relative ms timer (so wouldn't be negative), but I needed to initialize this to something less than zero so that alarm ranking logic will be forced to rank a brand new alarm (triggered 0 ms ago).

What's the significance of changing to U32? Is this being used as a timer?

What's the significance of changing to U32? Is this being used as a timer?

DEN-14999: Fixed alarm mgmt issue where top alarm may not be set if LOW priority alarm triggered. Ensure HD pump on flag is set to FALSE when pump controller is in off state.

  1. … 1 more file in changeset.
Should venPres < venLowLimit be done outside if statement?

Should venPres < venLowLimit be done outside if statement?

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14906-dialyzer-de-prime' into develop

DEN-14906: Increased HD pump max current persistence from 2 to 5 seconds. Added 1 minute delay following start/resume of dialysis where venous pressure limits are set to max selectable to allow pumps to stabilize.

  1. … 2 more files in changeset.
HD-DEN-14557_Dvt 012 HD Valve Transition Time Out Valve Position Out OF Target During Cartridge
HD-DEN-14557_Dvt 012 HD Valve Transition Time Out Valve Position Out OF Target During Cartridge
DEN-14557 updated the code set the valves status to position unknown if the tranistion to a position timedout.

DEN-14906: Fix response to Dialin requiring ACK when it should not.

  1. … 1 more file in changeset.
Merge branch 'staging' into DEN-14830-incorrect-dialysate-temperature-alarm-triggered

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.