Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15023: reworked syringe empty check to consider force and hw variance. Added t_par check for maximum heparin. renamed syringe occlusion and gave occlusion alarm persistence outside of priming

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15024-fw-srshd-1333-vtchd-283-alarm_id_hd_syringe_pump_not_stopped_error-id-111-does-not' into develop

Add testStartTime, audioTestStartTime, alarmAudioTestToneRequested and to header

Add testStartTime, audioTestStartTime, alarmAudioTestToneRequested and to header

Extra blank line.

Extra blank line.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14325 addressed the code review comments

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

It is not needed. So I removed it.

It is not needed. So I removed it.

Needed?

Needed?

Fixed.

Fixed.

Fixed.

Fixed.

Merge branch 'DEN-15183-in-treatment-rinseback-workflow-changes' into develop

  1. … 1 more file in changeset.
DEN-15189 Add activateSafetyShutdown to alarm ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE (62) if position C

Add treatmentTimeBroadcastTimerCtr to header.

Add treatmentTimeBroadcastTimerCtr to header.

DEN-15024 Change syringe pump off error count to max time windowed count.

Add doorOpened to header.

Add doorOpened to header.

HD-DEN-15189_Safety Shutdown Not Activating When HD Valve Position Out OF Range Alarm 62 Occurs
HD-DEN-15189_Safety Shutdown Not Activating When HD Valve Position Out OF Range Alarm 62 Occurs
DEN-15183: Removed pressure limit alarm exemption for pump ramping. Address sequence issue with invert dialyzer logic DEN-15195.

  1. … 1 more file in changeset.
HD-DEN-14325_FW Blood Leak Alarm Triggers When Cartridge Lever IS Unlocked
HD-DEN-14325_FW Blood Leak Alarm Triggers When Cartridge Lever IS Unlocked
DEN-15024 Fix order of saving last syringe position and syringe detection.

Merged DEN-15183 to develop.

  1. … 3 more files in changeset.
DEN-14325 added conditions to not check for blood leak alarm, air trap illegal level alarm, and the bubble detection alarm and self test request if the pump track is open