Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-15199 renamed the test configuration enums

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 10 more files in changeset.
RESOLVED in CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH

HD Staging Unreviewed Commits
HD Staging Unreviewed Commits
HD Staging Unreviewed Commits
HD Staging Unreviewed Commits
HD Staging Unreviewed Commits
HD Staging Unreviewed Commits
Is this needed? Why would a worn cartridge fail pressure test?

Is this needed? Why would a worn cartridge fail pressure test?

Why isn't this just an else? And if else should be swapped so normal action (alarm) comes first with if and the else should be the exception (else).

Why isn't this just an else? And if else should be swapped so normal action (alarm) comes first with if and the else should be the exception (else).

I don't think a worn cartridge should fail here (should still be dry). I think only wet cartridge should be checked here.

I don't think a worn cartridge should fail here (should still be dry). I think only wet cartridge should be checked here.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Recommend being more direct in naming: TEST_CONFIG_USE_WET_CARTRIDGE TEST_CONFIG_USE_WORN_CARTRIDGE

Recommend being more direct in naming:
TEST_CONFIG_USE_WET_CARTRIDGE
TEST_CONFIG_USE_WORN_CARTRIDGE

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Remove extra blank line.

Remove extra blank line.

DEN-15267: Refactored bubble detector driver to allow modes to push enable/disable instead of determining based on current op mode. Added support for new Dialin command to enable/disable venous bubble alarm detection.

  1. … 10 more files in changeset.
HD-DEN-15267_Request Create Dialin Command TO Enable Air Bubble Alarm IN Service Mode Air Bubble
HD-DEN-15267_Request Create Dialin Command TO Enable Air Bubble Alarm IN Service Mode Air Bubble
DEN-15104 Update for FE and OE errors. Minor correction to bloodLeakEmbModeCmd array. Enable venous pressure alarm which was commented out.

  1. … 1 more file in changeset.
Fixed merge conflicts from DEN-14262.

Please see changes.

Please see changes.

Done

Done

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-14262: Added a/c power loss flag to alarm info broadcast.

  1. … 2 more files in changeset.
DEN-15199 work in progress continued implementing the test configurations code

  1. … 6 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH