Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Add bubbleClearEnded.

Add bubbleClearEnded.

Make 2nd condition similar to 1st ( != TRUE ) so any value that is not 1 will take normal path. And should this be an OR (not AND)?

Make 2nd condition similar to 1st ( != TRUE ) so any value that is not 1 will take normal path. And should this be an OR (not AND)?

Blank line between declaration and code.

Blank line between declaration and code.

Blank line between declaration and code.

Blank line between declaration and code.

Blank line between declarations and code.

Blank line between declarations and code.

Add disinfectCancelReqID to header. Also argument stop.

Add disinfectCancelReqID to header. Also argument stop.

"bloodLeakPersitentCounter" should be bloodLeakPersistentCounter

"bloodLeakPersitentCounter" should be bloodLeakPersistentCounter

Add function header.

Add function header.

Add function header.

Add function header.

Add disinfectCancelReqID, currentStandbyState, currentDGCleaningMode

Add disinfectCancelReqID, currentStandbyState, currentDGCleaningMode

Isn't currentStandbyState an output?

Isn't currentStandbyState an output?

Add disinfectCancelReqID

Add disinfectCancelReqID

HD-DEN-15278_Blood Stop Air Trap Alarm Declares During Saline Bolus Treatment
HD-DEN-15278_Blood Stop Air Trap Alarm Declares During Saline Bolus Treatment
HD-DEN-15314_Usability Update Alarm Text TO Not Use Acronyms
HD-DEN-15314_Usability Update Alarm Text TO Not Use Acronyms
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

HD-DEN-15347_Usability Pressure Alarms During Recirculate Recirc State
HD-DEN-15347_Usability Pressure Alarms During Recirculate Recirc State
Done.

Done.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.