Add endTreatmentRequested

Add endTreatmentRequested

Add endTreatmentAlarmResponseRequest and sendLastTreatmentPeriodicData.

Add endTreatmentAlarmResponseRequest and sendLastTreatmentPeriodicData.

HD-DEN-15536_FW Usability Dry Self Tests Did Not Repeat After Dry Self Test Failure
HD-DEN-15536_FW Usability Dry Self Tests Did Not Repeat After Dry Self Test Failure
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done.

Done.

Add resumeRinsebackRequested.

Add resumeRinsebackRequested.

HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
HD-DEN-14934_Rinseback Back TO Treatment Workflow Missing Patient Reconnection Instructions
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done

Done

Done.

Done.

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

Add occlusionBaselineWasSet, selfTestStartTime, selfTestsResumeRequested (should be selfTestResumeRequested?)

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

You are stopping trimmer heater. Since DPi is running, you need to start the trimmer heater.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

See the "also" in my comment above.

See the "also" in my comment above.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Done

Done

Correct. Alarm 126 is not blood and dialysate recirculation.

Correct. Alarm 126 is not blood and dialysate recirculation.

Done

Done

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm? Also, since dialysate pump is on, need to start trimmer heater.

We are assuming blood pump was already stopped and VBA/VBV were closed by first blood leak alarm?

Also, since dialysate pump is on, need to start trimmer heater.