Merge remote-tracking branch 'origin/DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2' into develop

DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2

Reversed working revisions, to try minimal change.

  1. … 1 more file in changeset.
Update header.

Update header.

Update header

Update header

Update header

Update header

DEN-15886: Added events for air trap fill and air pump on/off events.

I reviewed DEN-15741 which is re: desire to stop DPi if user stopped with button. Everything I see changed in this module seems like a very complicated way to achieve that using properties that I d...

I reviewed DEN-15741 which is re: desire to stop DPi if user stopped with button.
Everything I see changed in this module seems like a very complicated way to achieve that using properties that I don't think were intended for this.
I think all we needed to do was check if user stop alarm is active and, in that special case, stop DPi instead of recirculating dialysate like it usually does in the recirc stop state.

Merge remote-tracking branch 'origin/DEN-15741-dpi-does-not-stop-when-recirculation-is-paused-2' into develop

Why was all of this necessary?

Why was all of this necessary?

Comment is copy. What is this flag for? Is it user requested?

Comment is copy. What is this flag for? Is it user requested?

DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2

Update to user stop state, and hard stop functions.

  1. … 1 more file in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

Done. dryPressureTestsCompleted is only set when test config is active so I don't want to include that in doxygen documentation.

Done. dryPressureTestsCompleted is only set when test config is active so I don't want to include that in doxygen documentation.

Add pressureSelfTestPreNormalStartTime, dryPressureTestsCompleted

Add pressureSelfTestPreNormalStartTime, dryPressureTestsCompleted

Add occlusionBaseLineWasSet

Add occlusionBaseLineWasSet

HD-DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2
HD-DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
HD-DEN-15893_Cartridge Lever IS Open Alarm When Cartridge Lever IS Closed
HD-DEN-15893_Cartridge Lever IS Open Alarm When Cartridge Lever IS Closed