Agreed. Done.

Agreed. Done.

Shouldn't this be calculated using the specific task interval?

Shouldn't this be calculated using the specific task interval?

HD-DENBUG-62_FW Alarm 130 When Changing QD Pressure
HD-DENBUG-62_FW Alarm 130 When Changing QD Pressure
Change simplified in TreatmentRecirc.c Pump control changes reverted. Christina's earlier testing confirmed these worked, similar to the findings in the DEN-16016 resolution.

Change simplified in TreatmentRecirc.c
Pump control changes reverted. Christina's earlier testing confirmed these worked, similar to the findings in the DEN-16016 resolution.

Change postponed for further testing. No change in BloodPumps.

Change postponed for further testing. No change in BloodPumps.

Approach simplified.

Approach simplified.

Update header.

Update header.

Update header

Update header

Update header

Update header

I reviewed DEN-15741 which is re: desire to stop DPi if user stopped with button. Everything I see changed in this module seems like a very complicated way to achieve that using properties that I d...

I reviewed DEN-15741 which is re: desire to stop DPi if user stopped with button.
Everything I see changed in this module seems like a very complicated way to achieve that using properties that I don't think were intended for this.
I think all we needed to do was check if user stop alarm is active and, in that special case, stop DPi instead of recirculating dialysate like it usually does in the recirc stop state.

Why was all of this necessary?

Why was all of this necessary?

Comment is copy. What is this flag for? Is it user requested?

Comment is copy. What is this flag for? Is it user requested?

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

Done. dryPressureTestsCompleted is only set when test config is active so I don't want to include that in doxygen documentation.

Done. dryPressureTestsCompleted is only set when test config is active so I don't want to include that in doxygen documentation.

Add pressureSelfTestPreNormalStartTime, dryPressureTestsCompleted

Add pressureSelfTestPreNormalStartTime, dryPressureTestsCompleted

Add occlusionBaseLineWasSet

Add occlusionBaseLineWasSet

HD-DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2
HD-DEN-15741_Dpi Does Not Stop When Recirculation IS Paused 2
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests