HD-DENBUG-101_Dpi Does Not Run When Blood Leak Sensor Fails TO Zero
HD-DENBUG-101_Dpi Does Not Run When Blood Leak Sensor Fails TO Zero
I think we need the minimum check too. uFRate can be negative if user sets volume to something lower than what has already been collected (negative volume) which would yield a negative rate which i...

I think we need the minimum check too. uFRate can be negative if user sets volume to something lower than what has already been collected (negative volume) which would yield a negative rate which is < 0.0.

Check casting. Result is U32 but could be assigned an S32.

Check casting. Result is U32 but could be assigned an S32.

HD-DEN-16446_HD Dev Test Clean UP Vectorcast
HD-DEN-16446_HD Dev Test Clean UP Vectorcast
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Thanks. Was not sure about that but wanted to cover it if bolus extended to treatment mode. Removed.

Thanks. Was not sure about that but wanted to cover it if bolus extended to treatment mode. Removed.

HD-DEN-16372_Srsdg 907 Not Met
HD-DEN-16372_Srsdg 907 Not Met
We should not be doing a Heparin bolus in treatment mode. It should be completed in pre-treatment mode. transitionToDialysis() function will resume Heparin dispense (not bolus) if necessary when tr...

We should not be doing a Heparin bolus in treatment mode. It should be completed in pre-treatment mode. transitionToDialysis() function will resume Heparin dispense (not bolus) if necessary when treatment is resumed.

HD-DENBUG-84_Heparin Bolus Does Not Resume After Treatment Pause
HD-DENBUG-84_Heparin Bolus Does Not Resume After Treatment Pause
HD-DENBUG-78_Dvt 21 Alarm 110 Triggers AT Start OF Treatment With Heparin ON
HD-DENBUG-78_Dvt 21 Alarm 110 Triggers AT Start OF Treatment With Heparin ON
HD-DENBUG-85_Treatment Log Missing Event For Heparin Paused BY User Request During Treatment
HD-DENBUG-85_Treatment Log Missing Event For Heparin Paused BY User Request During Treatment
HD-DEN-16456_Dpo Does Not Home IN Startup OR End Treatment
HD-DEN-16456_Dpo Does Not Home IN Startup OR End Treatment