Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
Fixed merge conflicts from local merge from staging.

  1. … 3 more files in changeset.
DEN-13154: Addressed code review comments.

Fixed merge conflict.

  1. … 1 more file in changeset.
Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

  1. … 3 more files in changeset.
more compiler fixes

DEN-14830: compiler fixes for develop.

DEN-14830: Implemented temperature recovery to be set at +-2 C of target temp. Implemented dialysate temperature alarms to have a clear period of 30s.

DEN-13154: Use raw FMD measured flow for alarm 312. Added capture of raw dialysate flow to facilitate this.

  1. … 1 more file in changeset.
DEN-13154: Add syringe pump stall check to preload and seek states.

Merge branch 'staging' into DEN-14763-dg_hd_dev-hd_dg_dvt-update-part-10

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-14834 updated the code to control the primary heater

  1. … 2 more files in changeset.
Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

  1. … 1 more file in changeset.
DEN-13154: Further fix to DEN-14814 re: do not require syringe removal to re-do preload operation.

  1. … 1 more file in changeset.
Fixed merge conflicts.

  1. … 1 more file in changeset.
DEN-13154: Removed dry self-test timeout alarm. Fixed bug DEN-14884.

  1. … 1 more file in changeset.
Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

DEN-13154: Removed obsolete pressure data from broadcast. Tightened no cartridge pressure range for PBA and PBo.

DEN-14830: reworked dialysate temp timeouts and conditions. Restored previous persistence alarms. added below 33 degC alarm

Merge branch 'staging' into DEN-14830-incorrect-dialysate-temperature-alarm-triggered

Merge branch 'DEN-14689-messaging-bugs' into develop

  1. … 3 more files in changeset.
DEN-14689: message naming sync

  1. … 10 more files in changeset.
Fixed merge conflicts.

  1. … 1 more file in changeset.
DEN-13154: Update to handling of insufficient Heparin alarm per DEN-14814 and DEN-14287.

  1. … 2 more files in changeset.
Fixed merge conflicts.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-14830: Fixed boolean logic for dialysate temperature alarms. Removed persistence counters for alarm 11 and 19 per PRS spec.

DEN-13154: Ensure alarm system has control of alarm lamp in fault mode in case fault occurred before lamp test completed. Do not request DG active alarms on startup if DG is not communicating.

  1. … 3 more files in changeset.
DEN-13154: Added prescribed flow rates to blood and dialysate flow broadcast messages per DEN-13152.