Controllers

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-5381: Removed test support macros

  1. … 2 more files in changeset.
DEN-4641: Updated persistent alarms to use ms instead of task interval counts.

  1. … 3 more files in changeset.
DEN-4641: Addressed code review feedback.

  1. … 3 more files in changeset.
DEN-4641: Merged with master. Fixed merge conflicts. Addressed code review feedback.

  1. … 17 more files in changeset.
DEN-4641: Addressed code review feedback.

  1. … 4 more files in changeset.
DEN-4641: Updates to UT.

  1. … 2 more files in changeset.
DEN-4641: Distributed remaining payload record definitions to associated modules and deleted MessagePayloads.h. Added messaging and verification of in-line pressure limit changes.

  1. … 10 more files in changeset.
DEN-4641: Fixed bug in last change.

DEN-4641: inline pressure range check only in treatment mode.

DEN-4641: Fixed safety line pull when pump running while off for dialysate outlet pump.

DEN-4640: Added inline pressure alarm checks.

  1. … 2 more files in changeset.
Fixed merge conflicts from master. Fixed some dev testing issues from merge.

  1. … 20 more files in changeset.
DEN-4641: Fixed PBA conversion formula.

DEN-4641: Pull safety line when cannot stop pumps.

DEN-4641: Changed HD pump Dialin set point override commands to simply set commands.

  1. … 1 more file in changeset.
DEN-4641: More fixes to HD pump motor speed measurement. Dialin cmd to clear all alarms will also restore op mode to prior if in fault mode.

  1. … 3 more files in changeset.
DEN-4641: Minor fixes to pump speed measurement and Dialin alarm features per verification.

  1. … 1 more file in changeset.
DEN-4641: Calculate HD pump motor speed more frequently. Added support for Dialin command to clear all alarms.

  1. … 7 more files in changeset.
DEN-4640: cosmetic changes.

DEN-4640: clear pump direction error counter if not in control state.

  1. … 1 more file in changeset.
DEN-4640: Added persistence to pump direction and blood pump max rotor speed alarms.

DEN-4640: Updated IT.

DEN-4640: Updated UT. Fixes per UT.

  1. … 1 more file in changeset.
DEN-4640: code review.

DEN-4640: Addressed code review comments.

  1. … 3 more files in changeset.
DEN-4641: Doxygen comment fixes.

DEN-4641: minor fixes.

  1. … 2 more files in changeset.
DEN-4641: Added pressure alarms. Changed occlusion data type to integer and unit from mmHg to unitless.

  1. … 3 more files in changeset.
DEN-4641: Updated inline pressure conversions.

  1. … 2 more files in changeset.
DEN-4641: Fixed PBA conversion to mmHg.