DEN-15203: code review comments.

  1. … 1 more file in changeset.
Merge branch 'DEN-15024-fw-srshd-1333-vtchd-283-alarm_id_hd_syringe_pump_not_stopped_error-id-111-does-not' into develop

DEN-15024 Increase syringe pump stop windowed error count to 10.

Merge branch 'DEN-15023-variation-in-syringe-pump-position-between-dvt-units' into develop

DEN-15023: reworded defines to handle advised fill volume and hw tolerance. removed positive position for force sensor.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-15183-2: Estimate initial DPo rate correction offset based on data collected by Systems team to improve UF control at treatment start.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15183-2 to develop.

  1. … 1 more file in changeset.
DEN-15203: Fixed alarm 32 to consider reverse direction with measured motor speed.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 8 more files in changeset.
Merge branch 'DEN-15183-in-treatment-rinseback-workflow-changes' into develop

DEN-15183: Fix for DEN-15198.

DEN-15023: reworked syringe empty check to consider force and hw variance. Added t_par check for maximum heparin. renamed syringe occlusion and gave occlusion alarm persistence outside of priming

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15024-fw-srshd-1333-vtchd-283-alarm_id_hd_syringe_pump_not_stopped_error-id-111-does-not' into develop

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14325 addressed the code review comments

Merge branch 'DEN-15183-in-treatment-rinseback-workflow-changes' into develop

  1. … 1 more file in changeset.
DEN-15189 Add activateSafetyShutdown to alarm ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE (62) if position C

DEN-15024 Change syringe pump off error count to max time windowed count.

DEN-15183: Removed pressure limit alarm exemption for pump ramping. Address sequence issue with invert dialyzer logic DEN-15195.

  1. … 1 more file in changeset.
DEN-15024 Fix order of saving last syringe position and syringe detection.

Merged DEN-15183 to develop.

  1. … 3 more files in changeset.
DEN-14325 added conditions to not check for blood leak alarm, air trap illegal level alarm, and the bubble detection alarm and self test request if the pump track is open

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 17 more files in changeset.
Merge branch 'staging' into DEN-15105-dialyzer-flip

DEN-15105: re-enabled a/c power loss alarm.

DEN-15024 Updated windowed timer param to new SRS requirement.

DEN-15105: Fixed DEN-14309.

  1. … 2 more files in changeset.