Updated the syringe pump retract function

  1. … 1 more file in changeset.
Merged DEN-13460 and resolved the conflicts

  1. … 1 more file in changeset.
Updated the drivers with the latest demo changes

  1. … 1 more file in changeset.
DEN-13460: Fixed syringe home issues.

  1. … 1 more file in changeset.
DEN-13460 updated the valves homing code

  1. … 2 more files in changeset.
DEN-13460: Relaxed alarm 159 threshold. Reverted back to old way of estimating implied motor speed from target flow for purposes of alarm 159.

DEN-13460: Fixed merge conflicts.

DEN-13460: Added persistence to pump direction error count alarm. Fixed comments and pos A average calc for valves.

  1. … 2 more files in changeset.
Merged DEN-13460

DEN-13460 updated the code for DVT

  1. … 2 more files in changeset.
DEN-13460: Fixed bug in syringe pump POST test.

DEN-13460: Baseline HD pump direction error counters after POST completes so we don't alarm due to power-up.

DEN-13460 added a config around the syringe pump self test

DEN-13460 added a config around the syringe pump self test

DEN-13460 reinstated the code that was removed in the HD valves

DEN-13460 reinstated the code that was removed in the HD valves

Merged DEN-13460

  1. … 5 more files in changeset.
DEN-13460 updated the code

Merge branch 'DEN-13460-dg_hd_dev-hd_dg_dvt-update-part-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-13460-dg_hd_dev-hd_dg_dvt-update-part-2

  1. … 1 more file in changeset.
DEN-13460 updated the code with DVT changes

  1. … 5 more files in changeset.
DEN-13460: Updated thresholds for wet transfer test. Updated valve home function to just go to position A if already homed.

  1. … 3 more files in changeset.
DEN-13460: Updated arterial pressure limits range. Added actuators homing to standby start state.

  1. … 3 more files in changeset.
DEN-12931 added isnan math function to prevent NaN values to be used in HD

DEN-12931: Updated blood flow estimation parameters for new tubing.

Merged staging and resolved the conflicts

  1. … 5 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 5 more files in changeset.
DEN-12931: Changed syringe pump calibration to expected DAC voltage. Added verification of calibration.

  1. … 1 more file in changeset.
DEN-12931: Refactored syringe detection alarm mgmt.

  1. … 5 more files in changeset.
DEN-12847 addressed code review comment

DEN-12847 merged staging and addressed VectorCAST issues

  1. … 4 more files in changeset.