Controllers

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-6402: Used enums from DG defs

  1. … 1 more file in changeset.
DEN-6402: Added DG valve setting change command

  1. … 6 more files in changeset.
DEN-6402: Stored command response for each possible command

  1. … 3 more files in changeset.
DEN-5887: Refactored load cell acceleration alarm. Tuned UF control.

  1. … 2 more files in changeset.
DEN-5887: Adjusted UF controller co-efficients.

DEN-5887: Increase load cell change threshold for non-treatment modes so will not cause alarm when load cell is tared. Remove obsolete messaging.

  1. … 5 more files in changeset.
DEN-6402: Removed home setup state and refactor big functions

  1. … 7 more files in changeset.
Reversed blood flow sensor output for V2 system.

DEN-6402: Fixed memcpy error and blood flow rate

  1. … 2 more files in changeset.
DEN-6372: Added over/under delivery checks/alarms. Added empty saline bag check.

  1. … 1 more file in changeset.
DEN-6402: Added PBo and PBa no cartridge self-test

DEN-6402: Added wet self-tests state machine

  1. … 5 more files in changeset.
DEN-6402: Added pressure occlusion sensors dry self test

  1. … 2 more files in changeset.
DEN-6402: Added flow meters and occlusion self tests

  1. … 2 more files in changeset.
DEN-5887: Pulled in addition of clear alarm condition functionality from other branch due to dependency. Added a little persistence to the load cell acceleration alarm.

  1. … 4 more files in changeset.
DEN-6453: Updated empty saline bag detection per feedback from Systems.

DEN-6200: Added clarification on handling the rest of reject reasons

DEN-5887: Added alarm for load cell acceleration.

  1. … 1 more file in changeset.
DEN-6453: Added empty saline bag check to saline bolus.

  1. … 2 more files in changeset.
DEN-6200: Added trimmer heater target temp to heater start/stop command

  1. … 2 more files in changeset.
DEN-6372: Added new modules for rinseback and re-circulate sub-modes of treatment mode. Fixed pre-treatment enum names per Quang change.

  1. … 10 more files in changeset.
DEN-6078 added an alarm after safety shutdown in HD valves

  1. … 3 more files in changeset.
DEN-6200: Added handler to DG command response

  1. … 4 more files in changeset.
DEN-6078 addressed a code review comment

DEN-6200: Handled pre-treatment alarm action

  1. … 1 more file in changeset.
DEN-6078 added safety shutdown to the valves driver

  1. … 5 more files in changeset.
DEN-6200: Reduced fill volume during priming

  1. … 3 more files in changeset.
DEN-6200: Updated persistent alarms

  1. … 1 more file in changeset.
DEN-6200: Removed unused functions and variables

  1. … 1 more file in changeset.
DEN-5674: Resolved merge conflicts from staging.

  1. … 13 more files in changeset.