Controllers

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-14999: Log DG command ID for rejected DG command alarm.

Fixed merge conflicts.

  1. … 6 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
DEN-14906: fixed merge conflicts from staging.

  1. … 8 more files in changeset.
DEN-14999: Fixed issue with repeated trimmer heater commands being sent to DG.

  1. … 1 more file in changeset.
Merge branch 'DEN-14906-dialyzer-de-prime' into develop

  1. … 1 more file in changeset.
DEN-14906: Fixed arterial high pressure check while in adjunct treatment states.

DEN-14999: Dont trigger alarm 312 unless in treatment mode or pre-treatment mode after priming complete.

DEN-14906: Addressed code review comment.

DEN-14999: Fixed alarm mgmt issue where top alarm may not be set if LOW priority alarm triggered. Ensure HD pump on flag is set to FALSE when pump controller is in off state.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-14906-dialyzer-de-prime' into develop

DEN-14906: Increased HD pump max current persistence from 2 to 5 seconds. Added 1 minute delay following start/resume of dialysis where venous pressure limits are set to max selectable to allow pumps to stabilize.

  1. … 2 more files in changeset.
DEN-14557 updated the code set the valves status to position unknown if the tranistion to a position timedout.

DEN-14906: Fix response to Dialin requiring ACK when it should not.

  1. … 1 more file in changeset.
Merge branch 'staging' into DEN-14830-incorrect-dialysate-temperature-alarm-triggered

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-14788: code review comments.

DEN-14999: Reduced BP occlusion clear threshold from 5500 to 5000 above baseline.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
DEN-14788: DIAL-242: added check for bubble sensor override.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 12 more files in changeset.
DEN-14906: Refactored DPo fixed PWM set function to work similar to DPi to better handle lower and negative PWM duty cycles.

DEN-14906: Fixed DPo fixed pwm function to allow reverse direction. Temporarily changed dialyzer re-prime interval to 5 minutes from 30 to speed up testing.

  1. … 1 more file in changeset.
Merge branch 'DEN-14906-dialyzer-de-prime' into develop

DEN-14906: In dialyzer re-prime, when running pumps in reverse, use fixed PWM functions to better control pumps.

  1. … 1 more file in changeset.
Merge branch 'staging' into DEN-14763-dg_hd_dev-hd_dg_dvt-update-part-10

  1. … 2 more files in changeset.
DEN-14763 added commands to be able to issue hard stop the HD pumps from Dialin

  1. … 4 more files in changeset.
DEN-14906: Exempt BLD alarm while dialyzer re-prime in progress.

  1. … 1 more file in changeset.
DEN-14830: Removed duplicate heater structure