DEN-15367 addressed the code review comments. Added code to check the payload of the message that is received from DG to indicate that DG is ready to dispense more water in RO permeate sample mode

  1. … 4 more files in changeset.
DEN-15558: Addressed code review comment.

Merge branch 'DEN-15558-fw-system-does-not-widen-venous-pressure-range-when-filling-air-trap-vbt-open-in' into develop

  1. … 1 more file in changeset.
DEN-15558: Extend exemption for venous low pressure alarm during air trap fill for 5 seconds after fill to allow pressure to recover.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15363-fw-handle-bld-sensor-communication-errors' into staging

DEN-15558: Reversed temporary change in dialyzer prime BP speed. Increased max venous pressure from 300 to 400 mmHg.

  1. … 1 more file in changeset.
Fixed merge conflict.

  1. … 1 more file in changeset.
Merged DEN-15459

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15459 changed the blood leak alarm detection to not trigger if the HD mode is not in treatment

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch 'staging' into DEN-15363-fw-handle-bld-sensor-communication-errors

DEN-15367 addressed the code review comments

  1. … 3 more files in changeset.
Merge branch 'DEN-15367-hd_dg_test_configuration_setup' into develop

  1. … 4 more files in changeset.
Fixed merge conflicts.

  1. … 1 more file in changeset.
DEN-15278: Added occlusion sensor value to no cartridge alarm data.

Merge branch 'DEN-15306-hd-fault-valve-position-alarm-after-restoring-ac-power' into develop

  1. … 3 more files in changeset.
DEN-15367 added a software fault to the blood leak embedded mode stat machine

  1. … 2 more files in changeset.
Merged DEN-15367

  1. … 4 more files in changeset.
DEN-15367, DEN-15252, DEN-15484, DEN-15485 added code to process the cleaning modes if they were started directly from DG. Added code to be able to start RO permeate sample from UI settings. Added the blood leak detector counter to the broadcast

  1. … 3 more files in changeset.
Fixed merge conflict.

  1. … 1 more file in changeset.
DEN-15278: Added rotor hall sensor state to BP, DPi and DPo broadcast messages.

DEN-15404, DEN-15252, DEN-15485 work in progress. Implementing code to handle cleaning modes in HD standby mode and removed the old code

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch 'DEN-15314-usability-update-alarm-text-to-not-use-acronyms' into develop

  1. … 1 more file in changeset.
DEN-15314: Added stall check to syringe prime operation. Extended dialysate bypass prime timeout and added a little histeresis to exit criteria for bypass prime state.

  1. … 1 more file in changeset.
Merged staging

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Fixed merge conflicts from DEN-15347.