DEN-15359: Misc fixes to art/ven pressure limit changes.

  1. … 1 more file in changeset.
Merge branch 'DEN-15359-usability-arterial-and-venous-pressure-alarm-limits' into develop

  1. … 1 more file in changeset.
DEN-15359: Fix s/w fault bug. Change so that dialysate flow change will also kick pressure limit s.m. back to stabilization state.

  1. … 1 more file in changeset.
DEN-15306-hd-fault-valve-position-alarm-after-restoring-ac-power

Corrected counter reset conditions.

Added CPLD.h

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 4 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merge branch 'DEN-15396-2-alarm-42-hd-fault-dpo-rotor-encountered-in-blood-prime-dvt-001' into staging

Merge branch 'DEN-15359-usability-arterial-and-venous-pressure-alarm-limits' into develop

  1. … 3 more files in changeset.
Merge branch 'DEN-13865-fw-duplicate-alarms-occur-during-post-treatment-rinseback-and-block-workflow' into develop

DEN-15359: Addressed code review comment.

DEN-15396 Code review comment, use macro.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15306-hd-fault-valve-position-alarm-after-restoring-ac-power

Made the power loss check uniform

isACPowerLoss is an alarm status, move to AlarmMgmt

Trigger with CPLD power loss flag, recover HW state when AC returns, start alarm block countdown after alarm is cleared.

  1. … 4 more files in changeset.
DEN-15359: handle blood flow rate change in middle of stabilization state.

DEN-15359: broadcast pressure data immediately after window change by user.

DEN-15359: Addressed code review comments.

  1. … 1 more file in changeset.
DEN-15359: finished arterial/venous pressure limits window update.

  1. … 3 more files in changeset.
DEN-15359: Additional implementation of arterial/venous pressure limit windows.

  1. … 3 more files in changeset.
DEN-15359: Added treatment parameter for asymmetric on venous pressure alarm limits.

  1. … 5 more files in changeset.
DEN-15359: Addressed DEN-15381.

  1. … 1 more file in changeset.
DEN-15359: Addressed DEN-15284.

DEN-15359: Updated tx params to change art/ven pressure limit params from min/max to window.

  1. … 4 more files in changeset.
DEN-13865: Addressed code review comment.

DEN-15396 Use greater of 10sec or 10 Rotor turn times for minimum persistent error time.

Merge branch 'DEN-13865-fw-duplicate-alarms-occur-during-post-treatment-rinseback-and-block-workflow' into develop

  1. … 1 more file in changeset.
DEN-13865: Implemented alarm system feature to prevent alarm re-trigger per new properties. Addressed DEN-15295.

  1. … 3 more files in changeset.
Merge remote-tracking branch 'origin/DEN-15363-fw-handle-bld-sensor-communication-errors' into develop

DEN-15363 Reuse alarm ID 63 for Blood Leak communications errors

Fixed merge conflict.

  1. … 3 more files in changeset.
DEN-15396-2: Fixed rotor speed chk persistence function to handle negative motor speeds.