DEN-15105: Started adding pre-tx patient connection sub-states.

  1. … 3 more files in changeset.
DEN-15105: Fix for DEN-15154 - saline bolus issue.

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 30 more files in changeset.
DEN-14999 updated the alarms table

  1. … 20 more files in changeset.
DEN-14999-2: Increased UF alarm thresholds to 100 mL for volume and 10 mL/min for rate.

DEN-14999 updated the code based on the alarms table

  1. … 2 more files in changeset.
DEN-14999-2: Removed alarm escalation feature.

  1. … 1 more file in changeset.
DEN-14999-2: Removed alarms 214, 215 and 217.

  1. … 2 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 15 more files in changeset.
DEN-14906: fixed merge conflicts from staging.

  1. … 11 more files in changeset.
DEN-14906: Fixed isues that arose when disabling re-prime feature.

  1. … 3 more files in changeset.
DEN-14906: Increased HD pump max current persistence from 2 to 5 seconds. Added 1 minute delay following start/resume of dialysis where venous pressure limits are set to max selectable to allow pumps to stabilize.

  1. … 6 more files in changeset.
DEN-14906: restored 30 min interval for dialyzer re-prime.

DEN-14906: Increased pump speed during dialyzer re-prime and reduce time in re-prime state in attempt to address over-current alarm.

DEN-14906: Added delay before resuming dialyzer re-prime to allow DPi to stop to prevent pressure spike.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-14888: Fix issue with initializing start/final reservoir volumes.

DEN-14906: Handle UF pause and auto-resume for dialyzer re-prime.

DEN-14906: Restored 30 minute interval for dialyzer re-prime. Reduced re-prime time from 40 to 35 seconds. Fixed so treatment time is paused during dialyzer re-prime.

  1. … 1 more file in changeset.
DEN-14906: Added dialyzer reprime state to treatment mode states broadcast message.

  1. … 3 more files in changeset.
DEN-14906: Fixed DPo fixed pwm function to allow reverse direction. Temporarily changed dialyzer re-prime interval to 5 minutes from 30 to speed up testing.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 11 more files in changeset.
DEN-14906: In dialyzer re-prime, when running pumps in reverse, use fixed PWM functions to better control pumps.

  1. … 2 more files in changeset.
DEN-14906: Fixed dialyzer re-prime transition from reverse to forward speed issue.

DEN-14888: Fixed dialyzer reprime transition bug.

DEN-14906: Addressed code review comments.

DEN-14906: Exempt BLD alarm while dialyzer re-prime in progress.

  1. … 1 more file in changeset.
DEN-14906-dialyzer-de-prime. Second round of code review changes

  1. … 1 more file in changeset.
DEN-14906-dialyzer-de-prime. Corrected a major error and cleaned up function headers

DEN-14906-dialyzer-de-prime. First round of addressing code review comments.

  1. … 1 more file in changeset.