DEN-4640: Fixes to treatment parameters response to validation request.

  1. … 4 more files in changeset.
DEN-5053: Added air trap alarms. Added actuator safe state enforcement to fault mode.

  1. … 7 more files in changeset.
DEN-5053: Fixed doxygen function headers for inputs/outputs details section.

  1. … 36 more files in changeset.
DEN-5053: Addressed code review feedback. Fixes to treatment workflow messaging with UI. Added an alarm lamp sync flag to alarm status message and increased transmission rate from 2 Hz to 4 Hz to allow alarm lamp sync flag to reach UI in timely manner. Added an assertion to give compiler error if new alarm is added to enumeration and not to alarm definition table. Update FPGA UT.

  1. … 22 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 60 more files in changeset.
DEN-4308 work in progress, added current relaxation to the valves module

  1. … 8 more files in changeset.
DEN-4211: Address code review feedback. Added additional UI message handling for treatment parameters.

  1. … 13 more files in changeset.
DEN-4211: Created saline bolus states.

  1. … 7 more files in changeset.
DEN-4211: Merge with doxygen fix branch.

  1. … 5 more files in changeset.
DEN-4534: More doxygen udpates.

  1. … 17 more files in changeset.
DEN-4534: Doxygen updates.

  1. … 38 more files in changeset.
DEN-4211: Added stop treatment logic when treatment duration is reached.

  1. … 6 more files in changeset.
DEN-4211: Updated format of @params in all modules.

  1. … 30 more files in changeset.
Matched DG arrangement and fix for CAN Tx. Added inlet conductivity alarm ID.

  1. … 5 more files in changeset.
Handle CAN and UART error/warning notifications. Differentiated HD from DG s/w fault. Updated some older module/function headers to latest format.

  1. … 24 more files in changeset.
Bamboo Commit: Updated the Copyright with the current year and replaced tabs with 4 spaces

  1. … 70 more files in changeset.
UT updates. Added UT for DGInterface. Misc fixes per UT.

  1. … 5 more files in changeset.
Added misc test/debug build flags. Added persistence to CAN buffer overflow faults so that we can try to recover from one-off spurts when a node is coming online.

  1. … 10 more files in changeset.
Fixed UF change verification logic. Fixed doxygen comments.

  1. … 3 more files in changeset.
Made reservoir target volumes public. Set all UF volumes to F32 type for consistency.

  1. … 5 more files in changeset.
Addressed code review feedback. Fixed treatment time minimum range to UI message.

  1. … 1 more file in changeset.
Fixed message payload sizing issue on UF option confirmation message handling.

  1. … 2 more files in changeset.
Changed max dialysate volume from 180 to 150 L. Split UF change response messages into 2 separate messages to prevent UI confusion.

  1. … 2 more files in changeset.
Added safety shutdown override command support. Zero treatment duration param in response to UF adjustment message if time adjustment should be disabled.

  1. … 7 more files in changeset.
Moved modes and sub-modes enums to common header. Added sub-mode to HD op mode broadcast. Added current UF rate to UF rate change response to UI.

  1. … 29 more files in changeset.
Misc UF fixes. Changed CAN from FIQ to IRQ to see if helps with intermittent lockup.

  1. … 10 more files in changeset.
Implemented some test code of UF demo support in Standby mode. Started a state machine for treatment mode reservoir management.

  1. … 4 more files in changeset.
Added minimum 1 hr treatment to settings adjustment range checks. Changed blood/dialysate inlet pump flow rate filter approach. Changed blood/dialysate inlet pump control interval to 10 sec.

  1. … 2 more files in changeset.
UT/IT updates.

  1. … 3 more files in changeset.
Fixes for eval board build for Behrouz.

  1. … 10 more files in changeset.